Memory (main unit) |
|
Max. Record
length |
Standard |
250 Mpts (1 CH)
10 Mpts/CH (16 CH) |
/M1 option |
1 Gpts (1CH)
50 Mpts/CH (16 CH) |
/M2 option |
2 Gpts (1CH)
100 Mpts/CH (16 CH) |
|
* Dependent on modules and no. of channels |
Channels |
16CH/Slot, 128CH/Unit |
(Maximum simultaneous display waveform is 64 waveforms x 4 screen selectable) |
Number of plug-in module slots |
8 total
Max 4 for 720210 modules
|
Time axis accuracy |
±0.005% |
Display—- |
Display: |
10.4-inch TFT color LCD monitor, 1024×768(XGA) |
Display res. of waveform display: |
selectable either 801×656 (normal waveform display) or
1001×656 (wide waveform display) |
Display format: |
Max 3 simultaneous displays available
In addition to main, 2 more waveforms available among zoom 1, zoom 2, XY1, XY2, FFT1, FFT2 (/G2 option), Vector (/G5 option), Bar graph (/G5 option) |
Functions |
Acquisition mode: |
Normal |
Normal waveform acquisition |
Envelope |
Maximum sample rate regardless of record time, holds peak value |
Averaging |
Average count 2 to 65536 (2n steps) |
Box (av.) |
Increase A/D resolution up to 4 bits (max 16 bits) |
Dual capture |
Main waveform
(low speed) |
Maximum sample rate |
100kS/s (roll mode region) |
Maximum record length |
1G point (/M2, 1CH) |
Capture waveform
(high speed) |
Maximum sample rate |
100MS/s |
Maximum record length |
500k point |
Performs data acquisition on the same waveform at 2 different sample rates |
Real Time HD Recording |
Maximum sample rate |
Maximum1MS/s (1CH used), 100kS/s (16CH used) depends on channel used |
(/HD0,/HD1 option) |
Capacity |
Depends on HDD vacant capacity |
Action |
Data can be stored in the hard disc at the same time of acquisition in accordance with trigger mode |
|
Roll mode: |
It is effective when the trigger mode is set to auto/auto level/single/ON start, and time axis is greater than 100ms/div. |
History memory: |
Maximum 5000 pages |
PC Card Interface |
Slots |
2 (front panel (1), rear panel (1)) |
Supported card |
GPIB card (National Instruments NI PCMCIA-GPIB card),
Flash ATA memory card (PC card TYPE II), CF card + adapter
card, and various hard disk type PC cards |
Storage |
SD card slot: |
Memory cards conforms to SD, SDHC |
USB memory: |
Mass storage device which conforms to USB Mass Storage
Class Ver.1.1 |
External HDD |
Hard disc conforms to eSATA, FAT32 |
Built-in HDD |
2.5 inch, 500GB, FAT32 |
Environment |
Power supply: |
100 to 120VAC/220 to 240VAC (automatic switching) — 50/60Hz |
Power consumption: |
Max 200 VA |
Operating Temperature: |
5C° to 40C° |
Dimensions: |
Approx. 355 mm (W) x 259 mm (H) x 202mm (D), excluding the
grip and projections |
Weight: |
Approx. 800 g |
With 17 unique plug-in module types, the DL850E can handle nearly any measurement task
The DL850E is backwards-compatible with all modules of its predecessor, the DL750. In addition, four new modules have been added to the lineup. Combine modules at will to measure anything from minute voltages to high-speed, high withstand voltages.
Input |
Input |
- High-speed Analog Voltage (6 modules)
- Temperature (4 modules)
- Strain (2 modules)
- Logic Input
|
- Analog Voltage,
Acceleration
- Frequency
- CAN Monitoring
- CAN LIN
|
[Image]
Capture high speed transients during long term recording using “Dual capture”
Never miss the waveforms you want to capture.
- High speed (100 MS/s), High resolution (12-bit), 1kV isolated measurements.
— Yokogawa’s isoPRO technology offers industry leading isolation performance at the highest speeds. The isoPRO core technology is designed with energy savings applications in mind. It gives you the performance needed to develop high efficiency inverters, which employ high voltages, large currents, and high operating speeds.
Using 720210 High-speed 100 MS/s 12-Bit Isolation Module
- Using high speed optical fiber-based transmissions, the module achieves high speed ADC clock and data isolation.
[Image]
Powerful Real-Time Calculations and Analysis Functions
Armed with a dedicated digital signal processor the ScopeCorder can perform mathematical calculations such as arithmetic operators with coefficients, integrals and differentials, and higher order equations on acquired measurement data. The results of these calculations are displayed during waveform capture in real-time. In addition to mathematical operators, steep digital filters can also be selected to isolate or trigger on the amplitude of certain frequency components
Trend calculations such as active power, power factor, integrated power and harmonics, using a dedicated Digital Signal Processor (DSP) that is able to calculate and display up to 125-types of electrical power related parameters in real-time. This enables the user to display raw waveform signals such as voltages and currents along with power calculated parameters and even the capability to trigger on all of them. Data updating rate up to 100kS/s. Trend waveforms of each order of harmonics, bar-graphs and vector displays can be displayed. Both RMS and Power analysis modes are available. Besides the powerful power calculations.
Advanced – Measurement Points – Up to 128 CH of voltage input and 128 bits of logic input
- 16-CH Voltage Input Module (scanner type) can measure at 10 kS/s sample rate even when using all 16 channels. With this module populating all 8 input module
slots, the DL850E performs 128-CH voltage measurements.
- The Logic Input Module supports everything from TTL levels, to high voltage contact closures at up to 10 MS/s. With eight logic modules, the DL850E can monitor and capture 128 bits of logic.
[Image]
Large (up to 2 GPoint with /M2 option) memory offers long duration measurement and two instantaneous zoom locations
— Comes standard with 250 MPoints of memory, expandable with 1or 2 GPoint options. Large capacity memory does not simply provide longer durations of measurement.
- High speed update
The DL850E Series DSO Models not only updates the display at high speed, but also include a function to recal up to 2000 screens worth of past waveforms.
- High-speed screen updating alone does not allow users to take full advantage of the digital oscilloscope. Rather, the ability to redisplay and analyzing individual waveforms unleash the digital oscilloscope’s full potential.
[Image]
Long Duration, Continuous Saving of Waveforms -Continuous Data Recording
- Measured data can be streamed directly to a built-in 160 GB hard disk (/HD1 option) or through the external HDD interface (/HD0 option). With long duration evaluation testing, measurements can be performed at 100 kS/s on 16 channels simultaneously for 10 hours
- Data being continuously recorded on the DL850E/DL850E’s built-in HDD or external HDD can be transferred to a PC without stopping measurement. You can display and analyze the transferred waveform data using Xviewer, an accessory program for the PC.
Search & Zoom – History Function
— When you spot an abnormal phenomenon during repetitive high speed measurements, often the anomaly has disappeared from the screen by the time you press Stop. Always active, the "History" function automatically divides the long memory into up to 5,000 "history waveforms" that can be re-displayed at any time.
- Searching history waveforms
When you want to extract specific c abnormal phenomena, you can perform condition-based searches inside the history waveforms. You can create a rectangular zone on screen and extract only waveforms that pass through or do not pass through the zone. You can also extract data based on parameters such as amplitude or RMS.
[Image]
Dual-window Zoom function simultaneously zooms in on two areas
- Two individual zoom factors and positions can be set with independent time scales and displayed simultaneously. Also, using the auto scroll function, you can automatically scroll waveforms captured in long memory and change the position of the zoom areas. Choose any display position with forward, backward, fast forward, pause, and other controls. Dual-window zoom function
[Image]
Catch transients in durability with high-speed sampling – Dual Capture
- To visualize long term trends in durability testing and other situations, data is typically acquired at low speed sample rates. On the other hand, suddenly-occurring transitional phenomena should be captured at high-speed sample rates.
The "Dual Capture" feature resolves these conflicting requirements by recording at two different sampling rates.
- Zoom Waveform
– You can record up to 5,000 phenomena of high speed trigger measurements (up to 100 MS/s) at a record length of 5-500 kPoints while taking trend measurements at up to 100 kS/s.
- Event Waveform
-Displays the timing at which high speed "sub" waveforms are acquired
- Low Speed Main Waveform
– Max: 100 kS/s Trend waveform displayed in a low-speed Roll mode
- Capture Waveform
– Max 100 MS/s Capture transients with high speed trigger measurement
[Image]
Processes Noise Rejection and Executes Power Computations in Real Time – Real time Math
- The DL850E is armed with a dedicated DSP (digital signal processor) for computations that enables between-channel math during waveform capture. These between-channel computations are powerful because they can be set up separately from filter computations. In addition to FIR, IIR, Gauss, and moving average filters, you can use arithmetic with coefficients, integrals and
differentials, and higher-order equations.
[Image]
GPS Interface (/C30 option)
- A GPS antenna can be directly connected to the ScopeCorder side panel. The instruments time clock and the sampling clock will be synchronized to the GPS clock.
User Defined Computation (/G2 option)
- The DL comes standard with arithmetic, time shift, FFT, and other computations that enable you to display waveforms with offsets and skew corrections. And with user defined computations (/G2 option), you can create equations using a combination of differentials and integrals, digital filters, and a wealth of other functions.
[Image]
Automatically extract waveform amplitude, frequency, and other parameters – Waveform parameter and statistical computation
- Extract and display up to 24 of the 26 available waveform parameters (amplitude, frequency, etc.) simultaneously. Menus can be shown as lists of easy-to-read icons.
Detect Abnormal Waveforms, Notify Users, and Determine Pass/Fail – Go/NO-GO Determination
- The DL can determine whether waveforms or computed values of waveform parameters meet (GO) or do not meet (NO-GO) conditions that are specified in advance. Upon judgment of the measured results, a pre-set action is performed and users are notified that an abnormal waveform was observed, along with the pass/fail determination. This is a very useful function for such things as studying signals from manufacturing lines of electronic devices and tracing abnormal phenomena
[Image]
Synchronize multiple units performing simultaneous measurements
Synchronized measurement across multiple DL850E units is made possible by inputting an IRIG time code signal.* The DL850E/DL850E’s internal clock is also synchronized (locked) to the IRIG signal. Therefore, timing comparisons are highly precise even when continuously recording over long periods of time.
- You can make periodic observations remotely by connecting commercially available GPS receivers that have IRIG output and using the Time Trigger function
[Image]
Check the relationship between hysteresis and phase
XY display function
- You can confirm the relationship between two signals using the X-Y display. This can be applied to measurements such as the phase angle of two sine waves. You can select four combinations on the X and Y axes, and therefore display multiple X-Y waveforms simultaneously and find relationships between them. Simultaneous observation of X-Y waveforms and normal T-Y waveforms (waveform display using voltage and time axes) is also possible.
- (1) Ensures security Simply remove the drive after measurement to protect data. Or, keep restricted data only at the measurement site.
- (2) Increases capacity If the external hard drive becomes full, you can simply switch to a new one (requires a restart).
- (3) Hi-speed data transfer A data can be transferred at high speed between a PC and a hard drive
- Computing dynamic BH characteristics of a magnetic substance On the DL850E you can measure voltage and current, then analyze hysteresis of magnetic flux density B and magnetic field strength H. Energy loss generated by magnetostriction can be evaluated by measuring dynamic BH characteristics
[Image]
Snapshots
- With the push of "SNAP SHOT" key, you can save a "snapshot" of the measured waveform (the waveform displayed on screen). The waveform remains saved even if you restart measurement, therefore you can easily compare the snapshot with any newly measured waveforms. Snapshots can also be saved and loaded as files.
[Image]
Web server
- The Web Server function displays the screen of any networked DL850E/DL850E on a PC via Ethernet. From this screen, you can remotely start or stop measurement, update the DL’s display, and take snapshots (capture images) of the screens.
Saving screen images and displaying thumbnails
- Screen images can be saved to a specified storage medium in PNG, JPEG, or BMP format. These screen images can be imported into reports or other PC-created documents.
[Image]
|
Volts |
- 720210- High-speed 100 MS/s 12-Bit Isolation Module
- 701250 – High-Speed 10 MS/s, 12-Bit Isolation Module
- 701251 – High-Speed 1 MS/s, 16-Bit Isolation Module
- 701255 – High-Speed 10 MS/s, 12-Bit Non-Isolation Module
- 701260 – High-Voltage 100 kS/s, 16-Bit Isolation Module
- 701220 – High-Speed 200 kS/s, 16-Bit Isolation Module
|
|
Temp. Volts |
- 701261 – Universal (Voltage/ Temp.) Module
- 701262 – Universal (Voltage/ Temp.) Module with AAF
- 701265 – Temperature, High Precision Voltage Isolation Module
|
|
Strain |
- 701270 – Strain Module (NDIS)
- 701271 – Strain Module (DSUB, Shunt-Cal)
|
|
Volts |
- 701275 – Acceleration/Voltage Module (with AAF)
|
|
Hz |
- 701280 – Frequency Module
|
|
Logic |
- 720230 – Logic input Module
|
|
CAN |
- 720240 – High Speed 100 kS/s, 16-bit CAN Module
|
|
|
Model |
Sample Rate |
Res. |
|
Ch |
Max. V |
Acc |
Volts |
High speed · High voltage · Isolated |
•720210 |
100 MS/s |
12-Bit |
20 MHz |
2 |
1000 V*2
200 V*3 |
±0.5% |
16CH voltage measurement (Scan-type) |
•720220 |
200kS/s |
16-Bit |
5 kHz |
2 |
42V*3 |
±0.3% |
High noise immunity |
• 701250 |
10 MS/s |
12-Bit |
3 MHz |
2 |
600 V*2
250 V*3 |
±0.5% |
High sensitivity range (1mV/div), low noise(±100 μV typ.), and high noise immunity |
• 701251 |
1 MS/s |
16-Bit |
300 kHz |
2 |
600 V*2
140 V*3 |
±0.25% |
Non-isolation version of model 701250 |
• 701255 |
10 MS/s |
12-Bit |
3 MHz |
2 |
600 V*4
250 V*3 |
±0.5% |
with RMS, and high noise immunity |
• 701260 |
100 kS/s |
16-Bit |
40 kHz |
2 |
1000 V*2
850 V*3 |
±0.25% |
Temp.
Volts |
Thermocouple (K, E, J, T, L, U, N, R, S, B, W, iron-doped gold/chromel) |
• 701261 |
100 kS/s (V),
500 S/s (Temp) |
16-Bit (V),
0.1: (Temp) |
40 kHz (V),
100 Hz (Temp) |
2 |
42 V |
±0.25% (V) |
Thermocouple (K, E, J, T, L, U, N, R, S, B, W, iron-doped gold/chromel), with AAF |
• 701262 |
100 kS/s (V),
500 S/s (Temp) |
>16-Bit (V),
0.1: (Temp) |
40 kHz (V),
100 Hz (Temp) |
2 |
42 V |
±0.25% (V) |
Thermocouple (K, E, J, T, L, U, N, R, S, B, W, iron-doped gold/chromel), high sensitivity range
(0.1mV/div), and low noise (±4 μV typ.) |
• 701265 |
500 S/s (V),
500 S/s (Temp) |
16-Bit (V),
0.1: (Temp) |
100 Hz |
2 |
42 V |
±0.08 (V) |
Strain |
Supports strain NDIS, 2, 5, 10 V built-in bridge power supply |
• 701270 |
100 kS/s |
16-Bit |
20 kHz |
2 |
10 V |
±0.5% (Strain) |
Supports strain DSUB, 2, 5, 10 V built-in bridge power supply, and shunt CAL |
• 701271 |
100 kS/s |
16-Bit |
20 kHz |
2 |
10 V |
±0.5% (Strain) |
Analog
Voltage
Accl. |
built-in anti-aliasing filter, Supports built-in amp type acceleration sensors
(4 mA/22 V) |
• 701275 |
100 kS/s |
16-Bit |
40 kHz |
2 |
42 V |
±0.25% (V)
±0.5% (Acc.) |
Hz |
Measurement frequency of 0.01 Hz to 200 kHz, Measured parameters (frequency, rpm, period,
duty, power supply frequency, distance, speed) |
• 701280 |
25 kS/s |
16-Bit |
res. 50 ns |
2 |
420 V*2
42 V*3 |
±0.1% (Hz) |
Logic |
8-bit/port) x 2, compatible with four-type of logic probe (sold separately) |
•720230 |
10 kS/s |
— |
— |
8-bit
x 2 ports |
— |
— |
CAN |
CAN Data of max. 16-bit allowable |
•720240 |
100 kS/s |
— |
— |
2 |
10V |
— |
*1: Probes are not included with any modules.
*2: In combination with 10:1 probe model 700929
*3: Direct input
*4: In combination with 10:1 probe model 701940 |
STANDARD ACCESSORIES |
|
- Passive Probes
- Isolated Probes
- Differential Probes
- Current Probes
- 100:1 Probes
- Bridge Head Probes
- Logic Probes
- Software
- Miscellaneous- Leads
|
OPTIONAL ACCESSORIES (Brochure Download) |
Passive Probes |
 |
701940 — 10 MΩ (10:1), 10 MHz, 1.5 m — Passive probe Passive Probe |
Isolated Probes |
 |
701947 — 100:1 DC to 200MHz (-3dB) /100MΩ/1,000V — Isolated probe 100:1 Probe |
Differential Probes |
 |
700924 — DC up to 100 MHz (-3 dB) bandwidth/4MΩ/max. — Differential Probe Differential probe capable of wide-band, high-voltage floating measurements |
 |
701926 —DC to 50MHz bandwidth / 50MΩ / max. ±5000 Vrms—Differential Probe High Voltage Differential Probe for the DL series |
Current Probes |
 |
701930 — DC to 10 MHz bandwidth, 150 Arms — Current probe Bandwidth: DC to 10 MHz (-3dB) |
 |
701931 —DC to 2 MHz bandwidth, 500 Arms — Current probe Bandwidth: DC to 2 MHz (-3dB) |
 |
701933 — DC to 50 MHz BW, 30 Arms — Current probe Bandwidth: DC to 2/10/50/100 MHz; current measurement: Max 30/150/500 A
1 MΩ BNC inputs without the need for an external amplifier |
100:1 Probes |
 |
701947 — 100:1 DC to 200MHz (-3dB) /100MΩ/1,000V — Isolated probe 100:1 Probe |
Bridge Head Probes |
 |
701955 —(NDIS, 120 Ω)—Bridge Head Probe Bridge resistance: 120Ω |
 |
701956 —(NDIS, 350Ω)—Bridge Head Probe Bridge resistance: 350Ω |
 |
701957 —(Dsub, 120 Ω)—Bridge Head Pro bee D-sub cable (5 m) included. Supports Shunt-Cal. |
 |
701958 —(Dsub, 120 Ω)—Bridge Head Probe D-sub cable (5 m) included. Supports Shunt-Cal. |
Logic Probes |
|
702911 — Logic Probe
8-Bit, 1 m, non-Isolated, TTL level/Contact Input, Includes one each of the B9879PX and B9879KX connection leads. |
|
702912 — Logic Probe
8-Bit, 3 m, non-Isolated, TTL level/Contact Input, Includes one each of the B9879PX and B9879KX connection leads. |
|
700986 — 100kΩ, ±42 V (DC +AC peak) 1.5m, Non-isolated— Logic Probe
8-bit, non-isolated, response speed: 1µs, two measurement leads (B9879PX and B9879KX) included |
|
700987 — 100kΩ, ±42 V (DC +AC peak) 1.5m, isolated— Logic Probe
8-Bit, each channel isolated |
Software |
 |
701992 —Xviewer— For DL series Software |
 |
701992-62E —Xviewer— For DL advanced series Software |
Miscellaneous |
 |
701934 PROBE POWER SUPPLY A power supply for current probes, FET probes, and differential probes. Supplies power for up to four probes, including large current probes. |
|
1:1 Safety BNC Adapter Lead(in combination with following) 1000 Vrms-CAT II (BNC-BNC)
701959— Safety Mini-Clip (Hook type) 1000 Vrms-CAT II, red and black set
701954— Large Alligator-Clip (Dolphin type) 1000 Vrms-CAT II, red and black set
758929— Alligator Clip Adaptor Set (Rated Voltage 1000 V) 1000 Vrms-CAT II, red and black set
758922— Alligator Clip Adaptor Set (Rated Voltage 300 V)300 Vrms-CAT II, red and black set
758321— (Fork Terminal Adapter Set) 1000 Vrms-CAT II, red and black set |
|
701902 —Safety BNC-BNC Cable (1 m) 1000 Vrms-CAT II (BNC-BNC) |
|
701903 —Safety BNC-BNC Cable (2 m) 1000 Vrms-CAT II (BNC-BNC) |
|
758924 — Safety BNC-banana Adapter 500 Vrms-CAT II |
|
758917 — Measurement Lead Set Measurement leads (2 per set) Alligator-Clip is required separately. |
|
720911 — External I/O Cable For external I/O connection |
|
B9988AE —Printer Roll Paper |
|
701963 — Soft Carrying Case for DL850E & DL850EVE |
|