Features
- Basic accuracy: 0.3°C (Typical of PT100)
- Corresponds to 14 types of RTD standard (JIS/IEC/GOST R)
- Sub display displays value of resistance
source and span%
- Corresponds to various types of source pattern
(Step sweep/ Linear sweep/ Manual step/ Span check)
- Corresponds to 2,3,4 wire. Realizes RTD simulation
- Measures output of RTD sensor as a thermometer
-
IIII [+] Read More
- Sources by 25% step automatically from 0% to 100% of span
- Sources by 25% step manually from 0% to 100% of span.
- Simple operation, easy to use
- Rotary switch range selector
- Linear Sweep Functon
- Lightweight, compact body
- Span Check Function
- Runs on 1.5 V AA-size Batteries or AC Adapter
Applications
- All types of loop test with control devices
- Valve test
|
- Temperature control, calibration
- Temperature sensor checking
|
Description
The CA300 Series is a high accuracy source and measure portable handheld calibrator used to inspect and maintain the accuracy of control devices such as temperature controllers, temperature transmitters, temperature sensors and other process devices which are used in process plants, industry machinery, and manufacturing. Available in 3 models, the CA310, CA320, and CA330 provide an ergonomic and friendly user-interface making it easy to use.
IIII [+] Read More
The CA330 is a dedicated process calibrator which is capable of performing inspections and calibrations of RTDs. It is two times accurate than the present CA12E with a high accuracy of ±0.3℃ error which is top class as a handy type to source resistance output and measure output of sensors. It not only corresponds to JIS standard and IEC standard thermocouples, but corresponds to GOST-R standard also. It corresponds to inspection and calibration with the summary of 14 types of RTD following the standards.
- Literature
- Specifications
|||| Click to View Source Specifications
Model: Yokogawa CA330 Calibrator |
RTD |
Coefficient |
Source/Meas. Temp. |
Source/Meas. Accuracy |
Excitation Current |
PT100 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.3V |
0.1-3mA |
PT100 |
3851 |
0.0‹C ≤ t ≤ +800.0‹C |
0.3+tx0.033% |
0.1-3mA |
PT100 |
3850 |
-200.0°C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
PT100 |
3850 |
0.0°C ≤ t ≤ +630.0°C |
0.3+tx0.033% |
0.1-3mA |
PT100 |
3916 |
-200.0°C ≤ < 0.0°C |
0.3 |
0.1-3mA |
PT100 |
3916 |
0.0°C ≤ t ≤ +510.0°C |
0.3+tx0.033% |
0.1-3mA |
PT100 |
3926 |
-200.0°C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
PT100 |
3926 |
0.0°C ≤ t ≤ +510.0°C |
0.3+tx0.033% |
0.1-3mA |
PT200 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.3 mV |
0.05-0.8mA |
PT200 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
0.3+tx0.050% |
0.05-0.8mA |
PT500 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.4 |
0.05-0.6mA |
PT500 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
0.4+tx0.033% |
0.05-0.6mA |
PT1000 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.2 |
0.05-0.6mA |
PT1000 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
0.2+tx0.033% |
0.05-0.6mA |
Cu10 |
427 |
-100.0°C ≤ t ≤ +260.0°C |
1.5 |
0.1-3mA |
Ni120 |
627 |
-80.0°C ≤ t ≤ +260.0°C |
0.2 |
0.1-3mA |
Extra RTD PT50 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
1 mV |
0.1-3mA |
Extra RTD PT50 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.4+tx0.050% |
0.1-3mA |
Extra RTD PT50G |
- |
-200.0°C ≤ t < 0.0°C |
0.4 |
0.1-3mA |
Extra RTD PT50G |
- |
0.0°C ≤ t ≤ +800.0°C |
0.4+tx0.050% |
0.1-3mA |
Extra RTD PT100G |
- |
-200.0°C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
Extra RTD PT100G |
- |
0.0°C ≤ t ≤ +630.0°C |
0.3+tx0.033% |
0.1-3mA |
Extra RTD Cu50M |
- |
-180.0°C ≤ t < 0.0°C |
0.4 |
0.1-3mA |
Extra RTD Cu50M |
- |
0.0°C ≤ t ≤ +200.0°C |
0.4+tx0.050% |
0.1-3mA |
Extra RTD Cu100M |
- |
-180°0‹C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
Extra RTD Cu100M |
- |
0.0°C ≤ t ≤ +200.0°C |
0.3+tx0.033% |
0.1-3mA |
Range |
Resolution |
Source & Meas. Range |
Source & Meas. Accuracy |
Note |
5000 Ω |
10 mΩ |
0.00 to 550.00Ω |
0.025% of setting/reading+0.1Ω |
0.1-3mA |
3000 Ω |
100 mΩ |
0.0 to 3300.0Ω |
0.025% of setting/reading+0.1Ω |
0.1-3mA |
|||| Click to View Measuring Specifications
Model: Yokogawa CA310 Calibrator |
RTD |
Coefficient |
Source/Meas. Temp. |
Source/Meas. Accuracy |
Excitation Current |
PT100 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.3V |
0.1-3mA |
PT100 |
3851 |
0.0‹C ≤ t ≤ +800.0‹C |
0.3+tx0.033% |
0.1-3mA |
PT100 |
3850 |
-200.0°C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
PT100 |
3850 |
0.0°C ≤ t ≤ +630.0°C |
0.3+tx0.033% |
0.1-3mA |
PT100 |
3916 |
-200.0°C ≤ < 0.0°C |
0.3 |
0.1-3mA |
PT100 |
3916 |
0.0°C ≤ t ≤ +510.0°C |
0.3+tx0.033% |
0.1-3mA |
PT100 |
3926 |
-200.0°C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
PT100 |
3926 |
0.0°C ≤ t ≤ +510.0°C |
0.3+tx0.033% |
0.1-3mA |
PT200 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.3 mV |
0.05-0.8mA |
PT200 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
0.3+tx0.050% |
0.05-0.8mA |
PT500 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.4 |
0.05-0.6mA |
PT500 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
0.4+tx0.033% |
0.05-0.6mA |
PT1000 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.2 |
0.05-0.6mA |
PT1000 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
0.2+tx0.033% |
0.05-0.6mA |
Cu10 |
427 |
-100.0°C ≤ t ≤ +260.0°C |
1.5 |
0.1-3mA |
Ni120 |
627 |
-80.0°C ≤ t ≤ +260.0°C |
0.2 |
0.1-3mA |
Extra RTD PT50 |
3851 |
0.0°C ≤ t ≤ +630.0°C |
1 mV |
0.1-3mA |
Extra RTD PT50 |
3851 |
-200.0°C ≤ t < 0.0°C |
0.4+tx0.050% |
0.1-3mA |
Extra RTD PT50G |
- |
-200.0°C ≤ t < 0.0°C |
0.4 |
0.1-3mA |
Extra RTD PT50G |
- |
0.0°C ≤ t ≤ +800.0°C |
0.4+tx0.050% |
0.1-3mA |
Extra RTD PT100G |
- |
-200.0°C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
Extra RTD PT100G |
- |
0.0°C ≤ t ≤ +630.0°C |
0.3+tx0.033% |
0.1-3mA |
Extra RTD Cu50M |
- |
-180.0°C ≤ t < 0.0°C |
0.4 |
0.1-3mA |
Extra RTD Cu50M |
- |
0.0°C ≤ t ≤ +200.0°C |
0.4+tx0.050% |
0.1-3mA |
Extra RTD Cu100M |
- |
-180°0‹C ≤ t < 0.0°C |
0.3 |
0.1-3mA |
Extra RTD Cu100M |
- |
0.0°C ≤ t ≤ +200.0°C |
0.3+tx0.033% |
0.1-3mA |
Range |
Resolution |
Source & Meas. Range |
Source & Meas. Accuracy |
Note |
5000 Ω |
10 mΩ |
0.00 to 550.00Ω |
0.025% of setting/reading+0.1Ω |
0.1-3mA |
3000 Ω |
100 mΩ |
0.0 to 3300.0Ω |
0.025% of setting/reading+0.1Ω |
0.1-3mA |
General Specifications |
Display: |
Segment LCD |
Backlight: |
LED (Selection of “Constantly ON”, “Constantly OFF” or “Auto off by approx. 2min”) |
Battery life: |
50 hours (5V source, load over 10kOhm), 25 hours
(20mA source, load under 5V)
|
Automatic Power-off: |
Approximately 20 minutes (Can be disabled by setting) |
Display Refresh Rate: |
Approx 1 sec. |
Level Setting
Response of generator: |
Approximately 1 second
(from when the output begins to change until when the
output level falls within the specified accuracy) |
Dimensions: |
Approximately Approx. 90 (W)×192 (H)×42 (D) |
Weight: |
Approximately 440 g |
Operating temperature/humidity: |
-10 to 55°C and 20 to 80%RH (no condensation) |
Storage temperature: |