Display |
Scaling: |
Customized to any scaling desired, from linear to specialized Cos Ø, square root distributions, with color applications for visual emphasis and to any desired engineering units. |
Specific accuracies |
Basic Accuracy: |
±2.0% of full scale basic accuracy class. |
Rectifier Type Meters: |
±3% of full scale basic accuracy class. |
Expanded Scale Voltmeters: |
0.5% of full scale |
Frequency Meters: |
± .15Hz @45-55Hz and 55-65Hz, ±0.093Hz @
58-62Hz, ±1.5Hz @350-450Hz |
Zero Suppressed Meters: |
Errors at 20% suppression = ±3%
of end-scale |
Higher Accuracy: |
1% of Full scale or better |
|
Position of use: |
Vertical (scale) standard
or to specified angle or position |
Full scale deflection angle: |
90 ° |
Movement: |
Taut Band |
The absence of friction
resulting from this permits greater sensitivities
and provides more rugged meters.Common
failures of conventional meters subjected to
shock testing are cracked jewels and/or
blunted pivots. |
Pivot and Jewel |
In this design the coil and pointer are
supported by polished steel pivots at each
end which fit into jewel bearings. Pivot and
jewel meters are recommended for high vibration
environments because of the
mechanism’s inherent stability and strength.
Accuracy and repeatability are almost equal
to the taut-band’s, and good performance
from the pivot and jewel construction is the
reason for its long popularity. |
Case: |
ABS plastic, Black with grey back |
Cover: |
Polycarbonate cover and window. |
Overload rating: |
Voltmeters (AC and DC) – 50%
momentary, 20% sustained. Ammeters
(AC and DC) momentary; 10 times the
rated current for 10 consecutive intervals
of .5 second with a 1 minute interval
between successive applications; sustained:
20% for six hours. |
Burden Data: |
AC Ammeter 5 amps @ 60 Hz, 0.5VA, Max. .5 Power factor lagging, AC voltmeter – unity power factor. |
Response time: |
3 seconds (Max.) – Microammeters
2 seconds (Max.) – All other DC ratings
2.5 seconds (Max.) – All AC ratings
Maximum 40% overshoot |
General: |
Operating Temp.: |
0 to 40°C (32 to 104°F) |
Storage Temp: |
-10 to 50°C (14 to 50°F) |
Extreme temperature range: |
-20°C to 65°C (-4 to 149°F)
Tropicalized or IP54 sealed units available |
Relative Humidity: |
95% (non condensing) |