• YOKOGAWA

Optical Time Domain Reflectometer AQ1200

(MFT-OTDR) Compact handheld multi-function chassis, fully equipped with powerful field testing functions & easy to operate

Yokogawa AQ1200 Optical Time Domain Reflectometer Series

Features

  • 7 models available with Wavelength capability from 1310nm (MMF) to 1650nm (SMF)
  • Tests up to 4 wavelengths
  • Wide Range of Modules Available
  • Enhanced support of FTTH to Metro, Core Network
  • High Dynamic Range (45dB)
  • Short Dead Zone (0.8 m)
  • Optional built in Dummy Fib
  • IIII [+] Read More
    • PON Measurement Capability
    • Auto and Multicore Loss measurement with LS & OPM interlock
    • Trace Analysis Functions
    • Visual fault location and fiber identification
    • Multi Fiber Measurement Function
    • Macro Bending Function
    • Remote Control Software
    • Video Fiber Inspection Probe
    • IPv4 PING

Applications

  • Installation and maintenance of optical fiber cables
  • Measure the trace in PON networks with up to 64 ports splitters

Description

the AQ1200 OTDR series multi-field testers are compact, lightweight, and moderately-priced, with all the functions needed in the field. All models support three different wavelengths, so that even a single unit can cover installation and maintenance work for an entire access network (a network between the station and subscribers). The AQ1205A, AQ1205E, and AQ1205F also feature a wide dynamic range and enhanced waveform quality, ensuring accurate measurements with fiber-optic communication networks that have a large-port-count optical splitting device with a high insertion loss.

IIII [+] Read More
The AQ1200 OTDR series are optimized for the installation and maintenance of optical fiber cables. Designed with ease of use in mind to simplify field testing, improve work efficiency and ensure qualify results. Seven models are offered, each with unique wavelength(s) based on their specific application.
The AQ1200 series testers all support passive optical network (PON) systems that branch out optical fibers from stations using an optical splitter and connect them to multiple subscribers. With PON systems, a measurement waveform tends to deform due to the attenuation of optical signals at a splitter. The AQ1200 OTDR series multi can determine whether any problem is caused by a line failure or signal attenuation, and can accurately pinpoint failure locations, if any.
 
 
  • Literature
  • Specifications
Literature and Technical Information
pdf icon

Optical Time Domain Reflectometers

++ AQ1200 Series Specification Brochure | 4 MB
++ AQ1200 Series User’s Guide | 14.5 MB
++ AQ1200 Communication Interface | 2.4 MB
Model: AQ1200 OTDR Series
Display: 5.7 inch color TFT LCD
Resolution: 640 (horizontal) × 480 (vertical) pixels
Models AQ1200A AQ1200B AQ1200C  
Wavelength: 1310±20(typ)
1550±20(typ)
1625±10 1650±5,1650±10  
Optical Port Fiber: PORT2  
Measured Fiber: SM(ITU-T G.652)  
Distance Range: 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 300, 400, 512  
Pulse Width: 3, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000  
Dynamic Range: 37 dB 38/36 dB  
Event Dead Zone: 0.75 m  
Attenuation Dead Zone: 4m/5m 7m  
Models AQ1200CE AQ1205A AQ1205E AQ1205F
Wavelength: 1310±20(typ)
1550±20(typ)
1625±10
1310±20(typ)
1550±20(typ)
1310±20(typ)
1550±20(typ)
1625±20(typ)
1310±20(typ)
1550±20(typ)
1650±5
1650±10
Applicable Fiber: PORT2,3 PORT2 PORT2, 3
Distance Range: 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 300, 400, 512
Pulse Width: 3, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000
Loss Meas. Accuracy: ±0.05dB or ±0.05dB/dB
Dynamic Range: 38/36,36 dB 42/40 dB 42/40/38 dB 42/40/37 dB
Event Dead Zone: 0.75 m
Attenuation Dead Zone: 4m/5m, 7m 4m/5m 4m/5m, 7m
Light source & optical power meter option
Power Meter Standard (/SLT) High power (/HLT) PON (/PPM)  
Wavelength: 850/1300/1310/1490/1550/1625/1650 nm or 800 to 1700 nm (1 nm steps) or CWDM wavelength (1270 to 1610 nm, 20 nm step) 1310/1490/1550 nm  
Power Range: CW: +10 to -70 dBm CW: +27 to -50 dBm CW: +10 to -70 dBm*1, +27 to -50 dBm  
CHOP: +7 to -60 dBm CHOP: +24 to -50 dBm ——
Noise Level: 0.5 nW (-63 dBm, 1310 nm) 50 nW (-43 dBm, 1310 nm) 0.5 nW (-63 dBm, 1310 nm),
50 nW (-43 dBm, 1550 nm)
 
Uncertainty: ±5% ±0.5 dB  
Readout Res.: 0.01  
Level Unit: Absolute: dBm, mW, μW, nW Relative: dB  
Modulation Mode: CW, CHOP (270 Hz/1 kHz/2 kHz)  
Average Function: 1, 10, 50, and 100 times  
Memory and Logging: Measurement data storage: 10 to 1000 data, Logging interval: 0.5, 1, 2, 5, or 10 sec.  
Auto Loss Test Function: Loss measurement with light source and optical power meter interlock ——–  
Light Source Standard (/SLT) High power (/HLT) PON (/PPM)  
Wavelength (nm): 1310/1550 ±25 nm (AQ1200A/E, AQ1205A/E/F), 1625 ±10 nm (AQ1200B/E), 1625 ±25 nm (AQ1205E),
1650 ±5 nm*5, 1650 ±10 nm*6 (AQ1200C, AQ1205F
 
Output Level (dBm): -3±1  
Level Stability (dB): ±0.05 (1310/1550nm), ±0.15 (1625/1650nm)  
Modulation Mode: CW, 270 Hz, 1 kHz, 2 kHz  
Fiber: SM (ITU-T G.652  
Specifications
Battery: Run time: 6 hours
Recharge time: 5 hours
Power : 100 to 240 VAC, 50/60 Hz
Temperature Range: 0 to 45°C (0 to 40°C when AC adapter is being used);
(0 to 35°C when battery is being charged)
Storage Temperature: –20 to 60 °C
Humidity Range: 20 to 85% RH (no condensation)
Dimensions: 217.5 (W) × 157 (H) × 74 (D) mm, excluding projections
Weight: Approx. 1 kg, including battery pack
Immunity: EN61326-1 Table 2 (for industrial locations)
Emission: EN61326-1 Class A, EN55011 Class A Group 1
Safety: EN61010-1