The quality of the power supply?
Wave window trigger
Instantaneous power failure Sag Impulse Trigger sensed Level window trigge Wave-window trigger

Want to monitor data remotely? Fax modem function With the OR100E
you can set up a wide-area data collection network over telephone lines using commercially available fax/modem PC cards. This
eliminates constraints on the physical distance between the oscillographic recorder and the monitoring station.

Temperature Measurement
In addition to the direct measurement capability (up to 500 Vrms), the new OR100E includes a temperature input adapter (788041-1) that works with type K thermocouples. The adapter is powered by the OR100E, so there’s no need for a special power supply when you use the recorder in the field or take long-duration measurements. The OR100E has a special temperature input adapter range (see the diagram on the right) for easy setup. (Screenshot)

Data Recorder Functions
using a PC card The OR100E lets you write measurements continuously and in real time to an ATA flash memory card (PCMCIA card). This means internal memory capacity is no longer a limitation on total recording time*1, allowing you to take long-duration measurements previously handled by data recorders. You can also display measured data stored in flash memory on the OR100E display for easy field checks.*2 In addition, ACRAWin32 is available to assist you in making reports when handling massive amounts of data.

*1 : Maximum sampling rate: 1.6 kS/sec;
*2 : Displays 128 kdata/ch steps

Four-mode power supply for field use
In addition to working with a 3.5-hour* rechargeable battery (788021), the OR100E/OR300E also can be powered by an AC adapter (100–240 V), DC adapters (12 V/24 V/48 V), and alkaline batteries. Choose the power supply type that best fits your needs.

Powerful Trigger Functions
The OR100ESeries provides powerful trigger functions for reliable measurement of monitored parameters. In Normal Trigger Mode, triggers are set for the rise and fall levels. Wave-window Trigger Mode allows you to monitor for power supply waveform abnormalities in real time. In addition, pre-trigger settings may be set as desired.

Sharp Display, Calculation Functions and High-Quality Recording Functions
Measurements are clearly displayed on a 5.7-inch backlit LCD display. You can view captured data and print out just the areas you need. Other capabilities include real-time recording, X-Y display and recording, and transient measurement (memory sampling when a trigger is sensed during real-time recording).

Caption_Header
Yokogawa
Model OR100E
Oscillographic Recorders (Standard Model)
 
previous URL Next URL
< Links >
 
 
 
 
OR100E
 
 
  Characteristics:  
 
  • Compact, lightweight and slim. Notebook size (B5) 4-channel isolated recorder
  • Fax/modem capability using a PC card
  • Powerful trigger functions
  • Four-mode power supply for field use
  • Standard external interfaces for flash ATA memory card and RS-232 communication
  • Temperature measurement using thermocouple
  • Data recorder function enabling extended-duration measurement
 
 
  Product Description:  
 
 

The OR100E series of Handy Oscillographic Recorders are complete multi-purpose recorders featuring up to four isolated analog input channels, fax/modem capability using a PC card, four-mode power supply and powerful trigger functions.

 
 
 
  General Information:  
space
    Measurement input
  • Input type: Floating unbalanced input, I/O isolation (channel independence)
  • Zero position: Can be moved within measurement range; null function included. Frequency characteristics (with filter off): DC to 40 kHz
    (+1/-3 dB, typical) Common mode rejection ratio (CMRR): 85 dB or greater (50/60 Hz; signal source resistance of 500 ( or less)
  • Low-pass filter: 5 Hz, 500 Hz, off
    Attenuation characteristic;-6 dB/octave
    Noise (with filter off, 10 mV/div range input shorted): 2.0 mVp-p (typical)
  • AD resolution: 12 bits (equivalent to 11bit internal processing resolution)
  • Maximum sampling speed: 400 kS/s (all channels simultaneously; 80 kS/s in wave-window)
  • Input impedance: 1 M ± 1%, 5 pF (at 40 kHz, typical)
  • Input terminal: Safety terminal (for banana plug)
  • Maximum input voltage and maximum floating voltage:
    Between H and L input terminals, between H-L input terminal and ground
  • Temperature input (with temperature input adapter)
  • Scale : Both upper and lower limits can be set in increments of 10*C.
  • Position : Can be moved in increments of 10*C.
  • Low-pass filter : 5 Hz (fixed)
  • Measurement range and accuracy : (at temperature of 23.5*C, following 30-minute warmup period)

    Memory function
  • Time axis: 200, 500 μs/div
    1, 2, 5, 10, 20, 50, 100, 200, 500 ms/div
    1, 2, 5, 10, 30 s/div
    1, 2 min/div
  • Time axis resolution: 80 points/div (measurement period is 1/80 of time axis)
  • Memory recording length:
    10, 20, 50, 100, 200, 400, 800, 1600, 3200*1, 6400*2 div
    *1: Only works on odd-numbered channel when two channels are connected together.
    *2: Only works on channel 1 when four channels are connected together.
  • Number of memory division blocks:
    32 maximum--varies depending on installed memory length, set memory length and whether or not wave window trigger is used.
  • Auto functions: Automatic printing, automatic statistical calculations, automatic saving (to external memory), automatic dialing (for faxing)

    Cursor functions:
  • One cursor: Measurements on all channels displayed simultaneously.
  • Two cursor: Time on all channels, as well as measurement differences or frequencies
  • Cursor calculation function: max/min/ave/rms for cursor range, surface area of cursor range Recording
  • Recording paper:Thermal paper roll (111 mm (width) x10 meters), effective recording
  • width: 104 mm
  • Paper feed accuracy: ±3%
  • Chart speeds: 2, 5, 10, 30 s/div; 1, 2, 5, 10, 30 min/div; 1 hour/div

    Recording formats:
  • T-Y recording: 4 analog channels + 8 logic bits; 1, 2, and 4 division recording capability (logic recording can be turned on/off separately for each bit)
  • Digital recording: measurements are recorded as digital values.
  • X-Y recording: X1-Y1, Y2, Y3. X axis is always 1 channel only.
    Recording size: 8 div x 8 div (80 mm x 80 mm)
  • Recording format options: Dots, lines
  • Recording length: 20 div, 200 div, 800 div, continuous
  • Recording line types: Three line thicknesses (analog waveforms) Printing function
  • Printed information: List (settings), scale (units), time print marker, chart speed, chart speed modification point marker, trigger sensing position, trigger time, trigger sensing channel, grid (thin line, baseline, off), channel number, TAG, etc.
  • Comments: Character string (20 characters per channel) or channel information printed in 100 mm intervals.
  • Channel number: Channel number or TAG name (7 characters per channel) printed on waveform. Real-time & memory
  • Description of operations: Normally memory sampling starts when trigger is detected during real-time recording. Normal Trigger
    Trigger sources: Analog channels 1-4, logic A and B, external trigger input, manual, timer
  • Trigger modes: Free, Single, Repeat Trigger combinations (conditions): AND/OR
  • Analog trigger types: Rise, fall, high, low, bi-slope,
    level window (in, out)
  • Trigger level setting: 1% FS increments

    Memory function
    Time axis: 200, 500 μs/div
    1, 2, 5, 10, 20, 50, 100, 200, 500 ms/div
    1, 2, 5, 10, 30 s/div
    1, 2 min/div
  • Time axis resolution: 80 points/div (measurement period is 1/80 of time axis)
  • Memory recording length:
    10, 20, 50, 100, 200, 400, 800, 1600, 3200*1, 6400*2 div
    *1: Only works on odd-numbered channel when two channels are connected together.
    *2: Only works on channel 1 when four channels are connected together.
  • Number of memory division blocks:
    32 maximum--varies depending on installed memory length, set memory length and whether or not wave window trigger is used. Auto functions: Automatic printing, automatic statistical calculations, automatic saving (to external memory), automatic dialing (for faxing)

    Wave-window trigger
  • Trigger modes: Single, repeat, free
  • Frequencies: 50 Hz, 60 Hz
  • Trigger combinations (conditions): OR on each analog channel

    Method for creating reference waveform:
  • Automatically generated from current input or specified parameters
  • Reference waveform parameters:
    Amplitude, tolerance, offset (1% increments for each), phase (in increments of 1°)
  • Trigger delay: -100% to 100% (in increments of 10%)
  • Sampling rates: 80 kS/s (1 ms/div), 40 kS/s (2 ms/div), 16 kS/s (5 ms/div), 8 kS/s(10 ms/div)
  • Memory length: Memory cannot be linked; maximum memory length for each channel is one-half that of normal triggers.
    Maximum memory length: 800 div
 
  Accuracy :  
space
Measurement range (V/div) Measurable range Accuracy
100mV 200mV ± 100.0mV ± (1% of FS + 1mV)
FS(20mV/div) ± 200.0mV ± (1% of FS + 1mV)
500mV FS(50mV/div) ± 500.0mV ± (1% of FS + 1mV)
1V FS(100mV/div) ± 1.000V ± (1% of FS + 1mV)
2V FS(200mV/div) ± 2.000V ± (1% of FS + 1mV)
5V FS(500mV/div) ± 5.000V ± (1% of FS + 1mV)
10V FS(1V /div) ± 10.00V ± (1% of FS + 1mV)
20V FS(2V /div) ± 20.00V ± (1% of FS + 1mV)
50V FS(5V /div) ± 50.00V ± (1% of FS + 1mV)
100V FS(10V /div) ± 100.0V ± (1% of FS + 1mV)
200V FS(20V /div) ± 200.0V ± (1% of FS + 1mV)
500V FS(50V /div) ± 500.0V ± (1% of FS + 1mV)
1000V FS(100V /div) ± 500.0V ± (1% of FS + 1mV)