Spacer Home Contact_us ob Opportunities ISO_Registration Privacy_Statement
 
Caption_Header
Yokogawa
Model FG200/FG300 Series
Signal Generators
 
previous URL Next URL
< Links >
 
 
 
 
FG200 / FG300
 
 
  Characteristics:  
 
  • Frequencies from 1 µHz to 15 MHz
  • Arbitrary sweep patterns (FG310/FG320)
  • Load and modify waveforms from other instruments (FG310/FG320)
  • Versatile sweep and modulation functions
  • External sweep control (/R1 option)
  • Amplitude and duty cycle
  • Multichannel output via synchronized operation
  • Up to 256 programmed sequence steps (FG310/FG320)
 
 
  Product Description:  
 
green_line
 
 

The FG200 Series function generators feature sweep and modulation capabilities with outstanding ease of operation. The FG300 Series adds arbitrary sweep and simple arbitrary waveform definition capabilities, plus sequencing functions, giving you high performance along with ease of use.
The powerful sweep and modulation capabilities of these instruments make them ideal for applications in mechatronics, and vehicle design and testing.

 
 
  Features :  
 
green_line
space
 
  • Frequencies from 1 µHz to 15 MHz
  • Frequency range:
    1 μHz to 15 MHz (Sine, Square)
    1 μHz to 200 kHz (Triangle, Pulse, Arbitrary waveform)
    High resolution:
    1 μHz or 9 digits max
  • Arbitrary sweep patterns (FG310/FG320)
    You define the waveforms by entering points within the scaled ranges for the X (time) and Y (parameter) axes. The waveform can be generated using linear, step, or spline interpolation between the points. Moreover, you can load the data in ASCII format from an MS-DOS floppy disko you can set up the data in Excel or the spreadsheet of your choice. You can use arbitrary waveforms thus defined as output waveforms or as sweep patterns.
  • Load and modify waveforms from other instruments (FG310/FG320)
    You can also use the floppy interface to load waveforms created with YOKOGAWA AG Series arbitrary waveform generators, or captured with YOKOGAWA digital oscilloscopes such as the DL4100, or DL5140/DL5180. You can then either output them as is, or, with the FG300 Series, make further modifications before use.
  • Versatile sweep and modulation functions
    With these instruments you can sweep not only frequency, but also other parameters as well, such as phase, amplitude, offset voltage, or duty cycle, with linear, log, linear step, log step, or arbitrary (FG310/FG320) sweep patterns. You can even sweep frequency and amplitude at the same time. Modulation capabilities include not only AM/FM, but also DSB-AM, phase modulation (PM), offset modulation, and PWM. The modulation waveform can be sine or triangle (with variable symmetry), pulse, (with variable duty cycle), or an arbitrary user-defined waveform (FG310/FG320).
  • External sweep control (/R1 option)
    Although these are digital function generators, you can still control the sweep parameters with an external analog signal. You can also control them directly with external digital signals. And you enjoy the high accuracy and repeatability that you can only get with digital technologies.
  • Amplitude and duty cycle
    With the FG200/FG300 Series, the output signal is not disrupted even if you continuously vary a parameter such as frequency, phase, amplitude, offset voltage, or duty cycle. Duty cycle setting range is 0% to 100%, with 0.01% resolution.
  • Multichannel output via synchronized operationBy linking multiple units in parallel with the accessory connecting cables, you can obtain three or even more channels of phase-synchronized signals, and even synchronize the sweep.
  • Up to 256 programmed sequence steps (FG310/FG320)
    The sequencing function lets you define sets of parameters for the frequency, phase, amplitude, offset voltage, and duty cycle and select them sequentially via trigger events. You can also specify sets of parameters directly with external digital control signals (with the /R1 option).
 
 
  Output and Operation Modes :  
 
green_line
space
 

WAVEFORM OUTPUT

  • Output channels FG220/FG320: 2 channels
    FG210/FG310: 1 channel
  • Output waveforms
    Sine, Square (fixed 50% duty cycle), Triangle (variable symmetry), Pulse (variable duty cycle), or Arbitrary waveform (FG310/FG320)
  • Output signal
    Continuous output(CONT):
    Waveform output continuously
    FG220/FG320: Switchable between phase-continuous and internal-channel synchronized output
    FG210/FG310: Phase-continuous only
    • Triggered output (TRIG)
      Output preset number (integer) of burst waveforms on trigger sync
    • Gated output (GATE)
      Output integer number of burst waveforms while enabled by gate DC output (DC) Output DC voltage

FREQUENCY

  • Frequency range
    • Sine or Square waveform                1 mHz to 15 MHz
    • Triangle or Pulse waveform             1 mHz to 200 kHz
    • Arbitrary waveform (FG310/FG320)  1 mHz to 200 kHz
  • Resolution 1 mHz or 9 digits max.
  • Frequency accuracy ±20 ppm
  • Frequency stability ±20 ppm (at ambient temperature of 5 to 40°C) Reference clock 40.2107 MHz

OUTPUT CHARACTERISTICS

  • Maximum output voltage***: ±10 V
  • Amplitude setting range***: ±20 Vpp (setting resolution 1 mVpp) (negative amplitude represents inverted waveform)
  • Amplitude accuracy***: ±(0.8% of setting + 14 mVrms)
    (1 kHz sine wave)
  • Amplitude frequency characteristics*
    • Sine waveform
      ≤100 kHz ±0.1 dB
      ≤1 MHz ±0.2 dB
      ≤10 MHz ±0.5 dB
      ≤15 MHz ±1 dB
    • Square/Pulse waveform (50% duty cycle)
      ≤10 kHz ±2%
    • Triangle waveform (50% symmetry)
      ≤10 kHz ±3%
  • Offset voltage setting range***: ±10 V (setting resolution 1 mV)
  • Offset voltage accuracy***: ±(0.3% of setting + 0.5% of amplitude setting +40 mV)
  • Output impedance: 50 W ±1% (open when output OFF)
  • DC output setting range***: ±10 V (setting resolution 1mV)
  • DC output accuracy***: ±(0.3% of setting + 20 mV)
  • Output attenuator setting range: 1/1, 1/10, 1/100
  • Output attenuator accuracy***: 0.2%
  • Interchannel crosstalk**(FG220/FG320): -65 dB max.

*   Rms measurements referenced to output at 1 kHz into
    50 W load, amplitude setting 20 Vpp, offset voltage 0V
**  Crosstalk between channels 1 and 2 with 50 W load, amplitude      setting 20 Vpp, offset voltage 0V, Ch1 set for 15 MHz sine wave, Ch2      set for 10 MHz sine wave.
*** High-impedance load

SINE WAVEFORM PURITY

  • Harmonics* (maximum of 2nd to 5th harmonic components)
    • 100 kHz -55 dBc max.
    • 1 MHz -45 dBc max.
    • 10 MHz -35 dBc max.
    • 15 MHz -25 dBc max.
  • Harmonic distortion* (rms value of 2nd to 5th harmonic components)
    100 kHz 0.3% max.
  • Spurious* (1 kHz to 100 MHz frequency range)
    100 kHz -55 dBc max.

* Measured with amplitude of 20 Vpp, offset voltage 0V, into 50 W load

SQUARE, PULSE AND TRIANGLE WAVEFORM CHARACTERISTICS

  • Rise time* Square waveform: 030 ns max. (10% to 90%)
  • Pulse waveform: 100 ns max. (10% to 90%)
  • Overshoot*: ±5% max. of output peak-to-peak value
  • Duty cycle setting*
    (Pulse waveform):
    • Setting range 0% to 100%
      (setting resolution 0.01% or 25 ns)
    • Time accuracy £10 kHz ±0.2% of (1/frequency setting)
    • Jitter 1 clock period
  • Symmetry* (Triangle waveform):
    • Setting range 0% to 100%
      (setting resolution 0.01% or 25 ns)
    • Jitter, 1 clock period

* Measured with amplitude of 20 Vpp, offset voltage 0V, into 50 W load

PHASE

  • Utilization: Starting/stopping phase of triggered or gated output. Channel-to-channel phase difference setting during 2-channel output
  • Setting range: -10000 deg to + 10000 deg (setting resolution 0.01 deg)

SWEEP CHARACTERISTICS

  • Sweep types:
    Linear, Log, Linear step, Log step, or Arbitrary pattern (FG310/FG320)
  • Sweepable parameters:
    Frequency, phase, amplitude, offset voltage, duty cycle, or simultaneous frequency and amplitude (sweepable range for each parameter is same as setting range during normal output) Sweep time setting range 1 ms to 10000 s (setting resolution 10 ms or 5 digits max.)
  • Sweep ratio: 0% to 100% (setting resolution 0.01% or 1.6 ms)
  • Sweep modes:
    • Continuous mode (REP)
      Perform continuously repeating sweep of each parameter
    • Single sweep (SINGLE)
      Perform single sweep synchronized to trigger
    • Single & hold sweep (SGL&HLD)
      Perform single sweep synchronized to trigger, continuing
      when finished to output waveform with final parameter
      value
 
 
  Downloads:  
green_line
 
  pdf_pic Downloads:General Specifications  
base

SYNCHRONIZED OPERATION

Number of units:
Up to eight units can be operated in synchronization

Output delay:
70 ns (typ.) for each unit
[25 ns (typ.) for each unit when triggered]

 

SEQUENCE (FG310/FG320)

Sequence mode
Sequential switching of output waveform sets of parameters under trigger control

Affected parameters
Can set frequency, phase, amplitude, offset voltage, and duty cycle for each step

Number of steps:
1 to 256 (returns to step 1 after last step)

 

ARBITRARY WAVEFORM (FG310/FG320)

Output amplitude resolution 12 bits

Memory length:
8192 points (not all points will be output if frequency exceeds 4.9 kHz)

Waveform definition functions:

Definable waveforms Output waveform, sweep pattern

Number of settings: 8

Interpolation
Linear, step, or spline

 

BUILT-IN FLOPPY DISK DRIVE (FG310/FG320)
Drive type:
3.5" floppy disk drive
Formats :
MS-DOS: 640 KB, 720 KB, 1.2 MB, and 1.44 MB

 

GP-IB COMMUNICATION INTERFACE

Electrical & mechanical specifications:
Conforms to IEEE St’d 488-1978

Functional specifications SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0

Protocol :Conforms to IEEE St’d 488.2-1987

 

GENERAL SPECIFICATIONS

Warm-up time:
30 minutes minimum

Operating temperature:
5°C to 40°C

Operating humidity range : 20% RH to 80% RH (max. wet bulb temperature 29°C, non-condensing)

Storage temperature:
-20°C to 60°C

Rated power voltage:
100 V AC to 240 V AC

Allowable range of power voltage variation:
90 V AC to 264 V AC

Rated power frequency:
50 Hz to 60 Hz

Allowable range of power frequency variation:
48 Hz to 63 Hz

Power consumption:
125 VA max.

Signal grounding:
Ground sides of all I/O connectors are connected to case ground

Dimensions: Approx.
213(W) ´ 132(H) ´ 350(D) mm (not including projections)

Weight: Approx. 5 kg (main unit only)

 

 

 
 