Full Set of communications to support Data Acquisition
Datum-Y supports standard protocols such as HTTP, FTP, SMTP
and SNTP based on Ethernet connection (10BASE-T, 100BASE-TX),
so no dedicated software is necessary.
You can also use the user authentication function to permit access
only to pre-registered users, in order to prevent illegal access to
Datum-Y from networks. Up to seven users (one administrator and
six users) can be pre-registered.
Web Server Function
You can monitor and operate Datum-Y’s screens
and functions via a web browser. The monitor
screen can be updated automatically (every 5, 10 or
30 seconds) or manually. Select the monitor page
where Datum-Y screens can be monitored, or the
operator page where both monitoring and operation
are possible. User authentication can be set for each
page to prevent illegal access.
FTP Server Function
You can output a list of files stored in Datum-Y’s internal
memory or connected external storage media, and
transfer, delete or otherwise manipulate desired files.
FTP Client Function
You can transfer the measurement data files/alarm
data files created in Datum-Y’s internal memory or
connected external storage media, to a FTP server.
Two file transfer destinations can be set (primary
and secondary). If the primary FTP server goes
down, files will be transferred to the secondary FTP
server
E-mail Delivery Function
You can send e-mail to notify occurrence/reset of
alarms, power recovery after outage, errors in
external storage media, FTP client errors, and
scheduled times. Up to two groups can be set, each
containing a different set of recipients.
RS485, RS232
Built-in Modbus Protocol Capability
Serial communication (RS-232, RS-485) can be
performed using a dedicated protocol as well as
Modbus RTU and Modbus ASCII protocols that
are supported by Datum-Y’s standard capability.
Datum-Y can be connected to other equipment
to acquire data simultaneously with
temperature /
voltage. 32 dedicated
communication channels are provided.
USB
Datum-Y connects
with a PC. You can acquire
data, change Datum-Y settings
or perform various other
operations on Datum-Y from a
PC.
Printer
RS-232 (Printer)
You can connect a printer
via RS-232 connection to
print paper reports on site. |
.jpg) |
|
|
|
| |
|
| |
 |
- World first in measuring instruments! Obtained IPv6 Ready Logo Phase-1
(ID:01-000273)
- Portable size (155×155×55mm) and lightweight (approx. 800g) lets you carry with one hand.
- Wide viewing angle TFT-LCD
- Best noise resistance performance in the category of handheld instruments.
- Improves resistance to impact by use of rubber boots (standard accessory).
- Web server, FTP server/client and e-mail distribution function are equipped as standard
- Auto set & Auto run function by inserting external memory.
- Alarm, Statistical operation and external function
|
|
|
| |
|
|
|
| |
| |
- Variety of simple temperature tests and evaluations (thermostatic chambers, electricfurnaces, inverter control units, etc.)
- Acquisition of facility maintenance data (data on the same time scale such as
temperature, voltage, pulse, etc.)
- Remote monitoring of analog data utilizing intranet LAN
- Acquisition of electric power and temperature data on the same time scale during
energy-saving operation
|
|
| |
| |
|
|
|
| |

|
| The Yokogawa Datum-Y portable data logger directly measures and acquires temperature (thermocouple/resistance temperature detector) and DC voltage. It provides two configurations of input terminals offering 8 and 16 channels, repectively individual channels can be set independently through universal insulated inputs. The Datum-Y is designed to achieve high performance and easy operation on site. Datum-Y provides wide ranging funtions and various communication capabilities to support diverse data acquisition applications. |
|
|
| |
| |
|
|
|
| |
| |
- All Channels Adopt Universal Insulated Inputs
The channels in the analog input part adopt insulated inputs, which means that
temperature (thermocouple/resistance temperature detector) and voltage can be
set differently for each channel. The digital inputs are equipped with one pulse
channel and two logic channels.
- Best-In-Class Noise Resistance
Datum-Y achieves excellent noise resistance rated among the best in the class of handy
measuring products. With Datum-Y, stable measurement is possible in various noise
environments, such as when measuring inverter-driven machines or thyrister-controlled units
that are commonly used for energy-saving purposes.
- Data Saving at 100ms
Datum-Y saves data at the maximum speed of 100 ms (when the 8-channel terminal block is
used).
You can save data in either binary format or ASCII format (for display in Excel*). Binary data
can be checked on site in the review mode or shown in waveforms using the supplied “D-TOOL” software.
- Acquisition of Large Amounts of Data
Using External Storage Media
Acquisition of Large Amounts of Data Using External
Storage Media
Datum-Y lets you save data not only in its internal
memory (16 MB), but also in external storage media
such as compact flash memory cards* and SD cards*
(up to 512 MB).
Data saved in these media can be copied to a USB
memory* for easy transfer to a PC.
|
|
| |
| |
|
|
|
| |
| |
- Spot Chech of Acquired Data
You can check measured data (binary) on Datum-Y right
away. Overall trend and alarm output can be checked on
the spot immediately after acquisition.
- While Datum-Y is logging data, you can display past data
and current data in the logging review mode for
comparison and identification of trend.
(This function is
available only when binary data is acquired.)
- Calculation Function
Datum-Y is capable of inter-channel calculations (among channels, between
channels and constants, etc.), statistical operations (minimum, maximum, average,
root-mean-square and peak values from start to end of logging), and scaling.
- Trigger Function
Datum-Y acquires data before and after a trigger (pre-trigger/trigger delay) to help you
detect and analyze the cause of an error. You can select the single mode (a trigger is issued
only once) or continuous mode (a trigger is issued every time the condition is satisfied).
- Alarm Function
Datum-Y can be fitted with up to four alarm output channels. One alarm can be set for
each input channel, and multiple channels can be combined freely with AND/OR
gates. You can also use the e-mail delivery function to notify specified e-mail
addresses upon occurrence or reset of an alarm
- Auto Set and Run Function
Simply connect an external storage medium (CF card, SD card or USB memory)
containing a predefined file (AUTORUN.SET), and Datum-Y will change the
necessary settings and take measurements automatically. (You need to enable the
automatic measurement function of Datum-Y beforehand.)
|
|
| |
| |
|
|
|
| |
| |
| ANALOG INPUT |
|
| Input method |
Floating unbalanced input, insulated between channels
(Terminal "b" is shared by resistance temperature detector inputs.) |
| Number of inputs |
8 channels (XL101), 16 channels (XL102, XL104) |
| Terminal shape |
Push-lock screw (XL101, XL102), M3 screw (XL104) |
Measurement
interval |
100 ms (only when the 8-channel terminal block is used), 200 ms, 500 ms, 1
sec, 2 sec, 5 sec, 10 sec, 20 sec, 30 sec, 1 min, 2 min, 5 min, 10 min, 20 min,
30 min, 1 hr |
| Input type |
TC (thermocouple), RTD (resistance temperature detector), DCV
(direct-current voltage)
* RTD for XL101 and XL102 only |
| Range and measurement range |
Reference operating conditions:
Temperature (23±2°C), humidity (55±10%RH),
power supply voltage (100 to 240 VAC), power supply frequency (50/60 Hz±1%
or less), warm-up (30 minutes or longer), without vibration, etc. that do not affect
the operation of the instrument |
| Input |
Range |
Measurement |
Accuracy |
Res. |
| DCV |
100mV |
-100.00 to 100.00mV |
±0.1% of f.s. |
10μV |
| 500mV |
-500.0 to 500.0mV |
100μV |
| 1V |
-1.0000 to 1.0000V |
100μV |
| 5V |
-5.000 to 5.000V |
1mV |
| 10V |
-10.000 to 10.000V |
1mV |
| 30V |
-30.00 to 30.00V |
10mV |
| 1-5V/f.s. |
1.000 to 5.000V |
1mV |
| TC |
R *1 |
0 to 1768°C |
±0.05% of f.s.±2°C *5 |
1°C |
| S *1 |
0 to 1768°C |
| B *1 |
600 to 1800°C |
| K *1 |
-200.0 to 1372.0°C |
±0.05% of f.s.±1°C *5 |
0.1°C |
| E *1 |
-200.0 to 1000.0°C |
| J *1 |
-200.0 to 1200.0°C |
| T *1 |
-200.0 to 400.0°C |
| N *1 |
-200.0 to 1300.0°C |
| W *2 |
0 to 2315°C |
±0.05% of f.s.±2°C *5 |
1°C |
| L *3 |
-200.0 to 900.0°C |
±0.05% of f.s.±1°C *5 |
0.1°C |
| U *3 |
-200.0 to 400.0°C |
| RTD |
Pt100 *4 |
-200.0 to 850.0°C |
±0.05% of f.s.±0.5°C *5 |
0.1°C |
| JPt100 *4 |
-200.0 to 500.0°C |
*1 R,S,B,K,E,J,T,N : IEC584-1 (1995), DIN IEC584, JIS C 1602-1995
*2 W : W-5% Rd/W-26% Rd (Hoskins Mfg. Co.), ASTM E988
*3 L : Fe-CuNi, DIN43710, U : Cu-CuNi, DIN43710
*4 Pt100 : JIS C 1604-1997, IEC 751-1995, DIN IEC751-1996
JPt100 : JIS C 1604-1989, JIS C 1606-1989
*5 "f.s." for TC and RTD means the full scale of the measuring range. |
|
| DIGITAL INPUT |
|
| Number of inputs |
Pulse input: 1 channel, Logic input: 2 channels |
| Input Specification |
Lo: Below 0.9 V or terminal short-circuited, Hi: 2.1 V or higher or
terminal open |
| Maximum Input |
10 VDC |
| Input |
Range |
Measurement |
Max. Res. |
Pulse
(Instant value) |
None |
50k/Measurement interval
0 to 50000c |
1c |
Pulse
(Integral Value) |
| 50kc/f.s. |
50k/Measurement interval |
1c |
| 500kc/f.s. |
10c |
| 5Mc/f.s. |
100c |
| 50Mc/f.s. |
1kc |
| 500Mc/f.s. |
10kc |
Pulse
(# of rev.) |
500rpm/f.s. |
50k/sec
(The number of pulses per
second is counted and
converted to the number
of revolution) |
- |
| 5krpm/f.s. |
- |
| 50krpm/f.s. |
- |
| 500krpm/f.s. |
- |
| Logic (DI) |
- |
- |
- |
|
| |
| CALCULATIONS |
|
Inter-channel
calculation |
Inter-channel calculation is possible.
(Measurement / calculation data / communication data, constant) |
| Linear scaling |
Ianges capable of scaling: DCV, TC, RTD, pulse
Available range of scaling: -30000 to 30000
Decimal point position: Selectable from 0.0000, 00.000, 000.00, 0000.0
and 00000
Unit: Can be set by the user (6 characters) or from standard list |
| Statistical operation |
Maximum value (MAX), minimum value (MIN), average value (AVE),
peak value (P-P) and root-mean-square value (RMS) between the
start and stop of logging |
| STORAGE |
|
| Internal memory: |
16 MB |
| External storage medium: |
Compact flash memory card (Type II), SD card, USB memory
(Only the copy function is supported by USB memory. Only those USB memories that
have been verified by YOKOGAWA are recommended.) |
| Storage data type |
|
| Type |
Description |
Format |
| Logging data |
Measurement is performed at specified intervals in logging
mode. / Instantaneous values (calculation data) are saved. |
Binary or ASCII |
| Manual sampling data |
Measurement is performed for all channels in free running
mode when a certain key is operated. / Calculation data
(instantaneous values) is saved. |
ASCII |
| Alarm data |
The same contents as the alarm summary are saved in
logging mode each time an alarm occurs. |
ASCII |
| Screen image data |
The image data of the currently displayed screen is saved
when a certain key is operated. |
BMP |
| Setting data |
The settings made to the instrument are saved when a
certain key is operated. |
Binary |
| Log data |
The same contents as the log display are saved when a
certain key is operated. |
ASCII |
| Backup file |
When data is not saved properly to the internal memory,
CF or SD card in logging mode (since, for instance, no card
has been inserted or the card is full), the data is saved to
the backup memory. |
Same format as
logging data |
|
| ALARM FUNCTIONS |
| Alarm type |
Hi (high limit), Lo (low limit), window-in (within specified upper/lower
range), window-out (outside specified upper/lower range)
(Only Hi and Lo are available for logic inputs.) |
| Alarm delay time |
Number of measurements: 0 to 36,000 |
| Display |
Alarm status is displayed in the status display area and measured
values are displayed in red when an alarm occurs.
(Selectable from non-hold and hold-type) |
| Hysteresis |
ON/OFF switchable (0.5% of span fixed, common to all channels)
4 channels (not insulated) |
| Buzzer |
ON/OFF switchable when being output |
| Recording |
Up to 120 sets of latest information can be recorded. |
| Output format |
Open collector, 5 V pull-up resistor (100 kΩ) |
| Contact capacity |
5 to 40 V, 100 mA |
| TRIGGER FUNCTIONS |
| Type |
Input to the external trigger input terminal, level (high limit, low limit, window-in,
window-out), alarm occurrence, time, timer (timer can only be used to stop logging)
For level, the trigger target channels must be specified. |
| Mode |
Single trigger (ends when the stop trigger is caused)
Continuous trigger (creates a file each time the trigger is caused) |
| Pre-trigger |
The data before the trigger is saved. |
| Trigger delay |
Data is saved when sampling has been performed the specified number
of times following the trigger. |
| POWER SUPPLY |
| AC POWER SUPPLY |
| Supply voltage |
100 to 240 VAC |
| Operating voltage |
90 to 132, 180 to 264 VAC |
| Frequency |
50/60 Hz |
| BATTERY |
| Battery used |
Dedicated lithium ion battery (2,400 mAh, 7.4 V) |
| Operation |
The battery can be charged on the main unit only.
The instrument runs on the AC adapter when both battery
and AC adapter are used. |
| Charging function |
The battery can be charged while the instrument is in use.
Charging takes approximately 8 hours. |
| Continuously operable
time |
Approx. 7 hours
(When used at 25°C, with measurement interval of 5
minutes or longer, backlight auto-off set to 5 minutes or less,
and no communication) |
| GENERAL
SPECIFICATIONS |
| Location for use: |
Indoor, at an altitude of 2000 meters or less |
| Operating Temp. |
0 to 50°C (0 to 40°C if a battery is used), |
| Humidity range: |
5 to 85%RH (no condensation) |
| Storage Temp. |
-20 to 60°C |
| Storage Humidity |
90%RH or less (no condensation) |
| INSULATION RESISTANCE |
| Terminals / Frame |
20 MΩ or higher (500 VDC) |
| Between input terminals** |
20 MΩ or higher (100 VDC) |
| Terminal and Digital I/O |
20 MΩ or higher (100 VDC) |
| WITHSTAND VOLTAGE |
| Terminals / Frame |
350 Vp-p (50/60 Hz), 1 min. |
| Between input terminals** |
350 Vp-p (50/60 Hz), 1 min. |
| Terminal and Digital I/O |
350 Vp-p (50/60 Hz), 1 min. |
| |
|
| Size |
Approx. 155 (W) x 155 (H) x 55 (D) mm
(Without projecting parts and rubber boot) |
| Weight |
Approx. 800 g (Without battery and rubber boot) |
| SAFETY
STANDARD |
| Safety Ratings |
EN61010-1
Measurement category I (circuit voltage used: ±60 VDC)
Pollution degree 2
Rated transient overvoltage 350 Vp-p |
| EMISSION
STANDARD |
| Complying standard: |
EN61326 Class A, EN55011 Class A Group 1
EN61000-3-2, EN61000-3-3
This product class A for use in an industrial environment and
may cause radio interference if used for domestic use.
Therefore, appropriate measures must be taken when using it
for domestic use. |
|
|
| |
| |
|
|
|
| |
| |
Specifications |
|
|
|
 |
|