| Input |
| Voltage: |
Voltage 115 / 230 VAC, ± 10 %, User selection |
| Frequency: |
50/60 Hz ± 5 % |
| Fuse: |
6.3 Amp 250V Slo-Blo |
| Dielectric Withstand Test Mode |
| Output Rating: |
5 KV @ 20mA AC
6 KV @ 5mA DC |
| Output Adjustment: |
Rating : 0 - 5 KV AC
0 - 6 KV DC
Resolution: 10 volt/step
Accuracy : ±(2% of setting + 5 volts) |
| Output Frequency: |
50 / 60 Hz selectable
|
| Wave Form: |
Sinewave, Distortion : <2 % THD |
| Dwell Time Setting: |
0 and 0.2 - 999.9 seconds, 0.1 second / step
“0” for continuous running |
| Failure Setting: |
High limit = 0.10 - 12.00 mA, 0.01 mA / step
Low limit = 0.00, 0.10 - 12.00 mA, 0.01 mA / step (0= OFF)
Accuracy = ± (2 % of setting + 0.02 mA) |
| Metering: |
Voltmeter (4 digits)
Range = AC 0.00 - 5.00 KV
Resolution = 0.01 KV
Accuracy = ± (2 % of reading + 10 V) |
DC mode
Range: 0.0 0 - 5.00 mA
Resolution: 0.01 mA/step
Accuracy: ± (2 % of setting + 0.02 m |
| Insulation Resistance Test Mode |
| Output Voltage: |
Range: 100 - 1000 Volts DC
Resolution: 1 volt/step
Accuracy: ± (2% of reading + 5 volts) |
| Output Voltage: |
Range: 0 - 1000 V
Resolution: 1 volt/step
Accuracy: ± (2% of reading + 2 counts) |
| Resistance Display |
| Range: |
1 - 1000 MΩ(4 Digit, Auto Ranging) |
| Resolution : |
MΩ |
500VDC
MΩ |
1000VDC
MΩ |
0.01
0.1 |
1.00 - 40.00
35.0 - 999.9 |
1.00 - 80.00
75.0 - 999.9 |
| Accuracy: |
± (3% of reading + 2 counts) at test voltage > 500V
± (7% of reading + 2 counts) at test voltage ≤ 500V |
| Resistance Display |
| HI Limit: |
Range: 0, 1 - 1000 MΩ (0 = Off) |
| LO-Limit: |
Range: 1 - 1000 MΩ |
| HI Limit: |
Range: 0, 1 - 1000 MΩ (0 = Off) |
| Delay Timer |
Range: 0, 0.5 - 999.9 sec (0 = Constant)
Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec) |
| Ground Bond Test Mode |
| Output Voltage: |
Range: AC 6 volts fixed |
| Output Frequency: |
Range: 60 or 50 Hz, User Selection
Accuracy: ± 0.1 % |
| Output Current : |
Range: 3.0 - 30.0 Amps AC
Resolution: 0.1 Amp/step
Accuracy: ± (2 % of Setting + 0.02 A) |
| Current Display : |
Range: 3.0 - 30.0 Amps AC
Resolution: 0.1 Amp/step
Accuracy: ± (2 % of Setting + 0.02 A) |
HI-Limit
LO-Limit |
Range: 0 - 510 mΩ
0 – 510 mΩ for 3.0 – 10.0 Amps
0 – 200 mΩ for 10.1 – 25.0 Amps
0 – 150 mΩ for 25.1 – 30.0 Amps
Resolution: 1 mΩ /step
Accuracy: ± (2 % of Setting + 2 mΩ) |
| Dwell Timer |
Range: 0, 0.5 - 999.9 sec (0 = Constant)
Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec) |
| Milliohm Offset |
Max. Offset Capability : 100 mΩ
Resolution: 1 mΩ / step
Accuracy: ± (2 % of Setting + 2 mΩ) |