Slaughter 2945 Series, AC/DC/IR Hipot Testers

 Characteristics:

green_line
 
  • Tamper proof front panel controls
  • Test setup memories
  • Low-current sense
  • Simple Menu Control
  • Electronic dwell settings
  • Electronic ramping
  • Front panel LCD
  • No load setup of trip current and output voltage
  • Storage of test program
  • Built-in Continuity test mode
  • Output voltage fine adjustment
  • Software calibration control
  • Current monitoring
  • Line and load regulation
  • PLC remote inputs and outputs
  • Flashing high voltage indicator
  • Insulation Resistance Test Mode
  • Interconnection Capability
 
 
  Product Description:  
space
green_line
 

space

The 2945 is a 3-in-1 Hipot tester that will perform AC, DC Hipot with a built-in Insulation Resistance tester. It will perform a 5 KV, 12 mA AC and a 6 KV, 10 mA DC Hipot test. It includes features such as a liquid crystal display, menu driven parameter control, electronic ramp, fully adjustable electronic dwell, 5 test set up memories, front panel lockout and a standard ground continuity check. The 2945 is also UL and TÜV safety agency listed.

 
 
  Specifications :  
 
green_line
 
Input
Voltage: 115 / 230V selectable, ± 15 % variation
47 - 63 Hz
Fuse: 115 VAC, 230VAC -- 3.15A fast acting 250VAC
Dielectric Withstand Test Mode
Output Voltage: Rating = AC 0 - 5000V, 2V / step, 12mA
DC 0 - 6000V, 2V / step, 5mA
Regulation = ± (1 % of output + 5V)
Voltage Setting: 0V - Max output rating, 10 volts/step Accuracy = ± (2 % of Setting + 5V)
(relative to displayed output)
Can be adjusted during operation via arrow keys.
Output Frequency: DC and 50 / 60 Hz selectable
Wave Form: Sinewave, Distortion : <2 % THD
Ripple: < 5 % at 6KVDC / 5mA
Dwell Time Setting: 0 and 0.2 - 999.9 seconds, 0.1 second / step
“0” for continuous running
Ramp Time Setting: 0 and 0.2 - 999.9 seconds, 0.1 second / step
0 ramp setting = 0.1 seconds
Failure Setting: AC mode
High limit = 0.10 - 12.00 mA, 0.01 mA / step
Low limit = 0.00, 0.10 - 12.00 mA, 0.01 mA / step (0= OFF)
Accuracy = ± (2 % of setting + 0.02 mA)
DC mode
Low limit = 0.02 - 5.00 mA, 0.01 mA / step
High limit = 0.00, 0.02 - 5.00 mA, 0.01 mA / step (0= OFF)
Accuracy = ± (2 % of setting + 0.02 mA)
Metering: Voltmeter (4 digits)
Range =
AC 0.00 - 5.00 KV
DC 0.00 - 6.00 KV
Resolution = 0.01 KV
Accuracy = ± (2 % of reading + 10 V)
Ammeter (4 digits)
Range = AC 0.10 - 12.00 mA
DC 0.02 - 5.00 mA
Resolution = 0.01 mA
Accuracy = ± (2 % of reading + 0.02mA)
Timer Display: Range = 0.0 - 999.9 seconds
Resolution = 0.1 second
Accuracy = ± (0.1 % of reading + 0.05 seconds)
Discharge Time: ≤300 ms
The maximum capacitive load vs output voltage:
0.20 µF --- < 1KV
0.10 µF --- < 2KV
0.06 µF --- < 3KV
0.050 µF --- < 4KV
0.040 µF --- < 5KV
0.015 µF --- < 6KV
Ground Continuity Check: Current = DC 0.1 A ± 0.01A, fixed
Max. ground resistance = 1Ω ± 0.1Ω, fixed
Insulation Resistance Test Mode
Output Voltage: Range = 100 - 1000 Volts DC
Resolution = 10 volt/step
Accuracy = ± (2% of reading + 5 volts)
Voltage Display: Range = 0 - 1000 V
Resolution = 10 volt/step
Accuracy = ± (2% of reading + 2 counts)
Resistance Display
Range: 1 - 1000 MΩ(4 Digit, Auto Ranging)
Resolution : 500VDC
1000VDC
0.01
0.1
1.00 - 40.00
35.0 - 999.9
1.00 - 80.00
75.0 - 999.9
Accuracy: ± (3% of reading + 2 counts) at test voltage > 500V
± (7% of reading + 2 counts) at test voltage ≤ 500V
High Resistance Limit Range = 0 - 1000 MΩ (0 = Off)
Low Resistance Limit Range = 1 - 1000 MΩ
Delay Timer Range = 0, 0.5 - 999.9 sec (0 = Constant)
Resolution = 0.1 sec/step
Accuracy = ± (0.1% + 0.05 sec)
 
 
  Downloads:  
space
green_line
 
  pdf_pic Downloads: General Specifications  
base

General Information
Remote Control and Signal Output
The following input and output signals are provided through the 9 pin D type connector:

Inputs:
Test, Reset
Outputs:
Pass, Fail, Processing

Memory
5 Memories.

Security
Key Lock capability to avoid unauthorized access to all test parameters. Memory Lock capability to avoid unauthorized access to memory locations.

Line Cord
Detachable 7 ft. (2.13m) power cable terminated in a three prong grounding plug.

Treminations
6ft. (1.83m) high voltage probe and 6 ft. high voltage lead. 6ft continuity lead with clip and 6ft return lead with clip.

Dimension
(W x H X D)
11 x 3.5 x 14.56 in.
(280 x 89 x 370 mm)

Weight 20 lbs (9 Kgs)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 