- Test setup memories.
- Low-current sense
- Simple Menu Control
- Electronic dwell settings
- Electronic ramping
- Front panel LCD
- No load setup of trip current and output voltage
- Storage of test program
- Built-in Continuity test mode
- Output voltage fine adjustment
- Software calibration control
- Current monitoring
- Line and load regulation
- PLC remote inputs and outputs
- Flashing high voltage indicator
|
|
|
| |
| |
|
|
|
| |
|

|
Featuring our most advanced technology, the 2925 is a 5 KV, 12 mA AC Hipot tester that allows manufacturers to quickly and cost-effectively perform AC Hipot tests. It includes features such as a liquid crystal display, menu driven parameter control, electronic ramp, fully adjustable electronic dwell, 5 test set up memories, front panel lockout and a standard ground continuity check. The 2925 is also UL and TÜV safety agency listed. |
|
|
| |
| |
|
|
| |
|
| |
|
| Input |
| Voltage: |
115 / 230V selectable, ± 10% variation
50 - 60 Hz ± 5% |
| Fuse: |
115 VAC, 230VAC -- 3.15A fast acting 250VAC |
| Output Voltage: |
Rating = AC 0 - 5000V, 2V / step, 12mA
Regulation = ± (1 % of output + 5V) |
| Voltage Setting: |
0V - Max output rating, 10 volts/step
Accuracy = ± (2 % of Setting + 5V) (relative to displayed output) Can be adjusted during operation via UP and DOWN arrow keys. |
| Output Frequency |
50 / 60 Hz selectable |
| Wave Form |
Sinewave, Distortion : <2 % THD |
| Dwell Time Setting: |
0 and 0.2 - 999.9 seconds, 0.1 second / step
“0” for continuous running |
| Ramp Time Setting: |
0 and 0.2 - 999.9 seconds, 0.1 second / step
0 ramp setting = 0.1 seconds |
| Failure Setting: |
High limit = 0.10 - 12.00 mA, 0.01 mA / step
Low limit = 0.00, 0.10 - 12.00 mA, 0.01 mA / step (0= OFF)
Accuracy = ± (2 % of setting + 0.02 mA) |
| Metering |
Voltmeter (4 digits)
Range = AC 0.00 - 5.00 KV
Resolution = 0.01 KV
Accuracy = ± (2 % of reading + 10 V) |
Ammeter (4 digits)
Range = AC 0.10 - 12.00 mA
Resolution = 0.01 mA
Accuracy = ± (2 % of reading + 0.02mA) |
| Timer Display: |
Range = 0.0 - 999.9 seconds
Resolution = 0.1 second
Accuracy = ± (0.1 % of reading + 0.05 seconds) |
| Ground Continuity Check: |
Current = DC 0.1 A ± 0.01A, fixed
Max. ground resistance = 1Ω ± 0.1Ω, fixed |
|
|
| |
| GENERAL SPECIFICATIONS |
| Remote Control |
The following input and output signals are
provided through the 9 pin D type connect
1. Remote control: Test and Reset
2. Outputs: Pass, Fail and Test in Process |
| Memory |
5 Memories |
| Security |
Key Lock capability to avoid unauthorized access
to all test parameters. Memory Lock capability to
avoid unauthorized access to memory locations.DC 0.00 - 6.00KV
Resolution: .01KV |
| Line Cord |
Detachable 7 ft (2.13 m) power cable terminated
in a three-prong grounding plug |
| Terminations |
6ft. (1.83m) high voltage probe and
6 ft. high
voltage lead.
6ft continuity lead with clip and 6ft
return lead with clip. |
| Environmental |
Operating Temperature: 32°? - 113°?F (0°? - 45°?C)
Relative Humidity - 0 to 95% |
| Calibration |
Traceable to National Institute of Standards and
Technology (NIST). Calibration controlled by
software. Adjustments are made through front panel
keypad in a restricted access calibration mode.
Calibration information stored in non-volatile memory. |
| Dimension |
(W x H x D) 11 x 3.5 x 14.56 in.
(280 x 89 x 370 mm) |
| Weight |
20 lbs. (9Kgs) |
|
| |
| |
|
|
|
|
| |
| |
General Specifications |
|
|
|
|
|
 |
|
|
|
|