- Tamper-Resistant Front Panel Design
- Simple Menu Control
- No load setup of trip current and output voltage
- Test Storage
- LED Display With Meter Memory
- Software calibration control
- Line and load regulation
- Flashing high voltage indicator
|
|
|
| |
| |
|
|
|
| |
|

|
The 1305 is 3 KV, 5 mA AC Hipot tester. This entry-level tester is designed with the basic features for simple and cost-effective production line AC Hipot testing. It features a LED display, menu driven parameter control, pulse mode, parameter storage and continuous or momentary dwell cycles. |
|
|
| |
| |
|
|
| |
|
| |
|
| Input |
| Voltage: |
115 / 230 selectable
+/- 15% variation, 47 – 63Hz |
| Fuse: |
115 VAC, 230 VAC - 2A fast acting 250 VAC |
| Output Voltage: |
Rating = 0 – 3000VAC
Regulation = +/- 1% of output + 5V |
| Voltage Setting: |
Rating = 0.01KV – 3.00KV, 10volts/step
Accuracy = +/- 2% of Setting + 5V
(relative to display output)
Can be adjusted during operation |
| Dwell Time Setting: |
On = Continuous
Off = Momentary |
| Pulse Mode: |
On = Automatic reset after failure
Off = Manual reset after failure |
| Failure Setting: |
High Limit = 0.00-5.00mA,0.01mA/step
Accuracy= +/- 2% of set + .02mA |
| Metering |
Voltmeter (3 digits)
Range = AC 0.00 – 3.00KV
Resolution = 0.01KV
Accuracy = +/- 2% of read +10V |
|
|
|
| |
| GENERAL SPECIFICATIONS |
| Remote Control |
Test start input through a 9-pin
D type connector. |
| Line Cord |
Detachable 7 ft (2.13 m) power cable terminated
in a three-prong grounding plug |
| Terminations |
HV Safety Retracting Probe with
6 ft. (1.82m) Lead.
Ground Return
Clip with 6 ft. (1.82m) Lead |
| Dimension |
(W x H X D) 4.75” x 5.75” x 14.5”
(120 x 146 x 370 mm) |
| Weight |
16.0 lbs. (7.25 Kgs) |
|
|
| |
| |
|
|
|
|
| |
| |
General Specifications |
|
|
|
|
|
 |
|
|
|
|