- Data logging capabilities
- Insulation measurement up to 200 Gohm
(BMM2580)
- 200 mA Continuity Range
- Voltage, current and kΩ resistance measurement
- Result storage and Data Logging
- PowerSuite Compatible RS232 output
- Download Manager Software included
- Direct Printer Output
- Electronic Locking Test Button
- Remote control switched probe
- Complete multimeter capability
- Frequency measurement
- mV transducer input
- Backlight
- Selectable locking test button
- Data storage and downloading
|
|
| |
| |
|
|
| |

|
The Megger BMM Series Multi-MEGGERS are a versatile range of hand-held insulation testers designed to extend the capability of insulation testers beyond anything available on the market today.
In addition to extremely high-sensitivity insulation resistance measurements (200 GΩ), these instruments offer complete multimeter testing capability and the ability to view the insulation measurement in terms of leakage current (µA). The top-of-the- range models also include data storage and download capability. The end user can now carry a single instrument to the test site rather than multiple instruments.
BMM2500: Top of the range contractor/service technician insulation tester and multimeter featuring result storage, printing, and downloading. Offers the same voltage ranges as the BMM2000.
BMM2580: Top of the range low voltage/specialty application insulation tester and multimeter. Features include result storage, printing, and downloading. Voltage ranges of 50, 100, 250, 500, and 1000 are also available in this model. Additionally, all units are supplied with the SP1 Remote Switch Probe as a standard accessory (at no additional cost) for hands-free operation. |
|
| |
| |
|
|
|
| |
|
- The instruments utilise a large backlit LCD display
incorporating a patented analog arc which incorporates
the benefits of electronic analog indication and
unambiguous digital readings. The analog scale
provides rapid identification of insulation condition
highlighting any variable readings and is complemented by
the precision and simplicity of the digital display.
The BMM2500 series provide the ideal mix of features for
the busy electrical contractor requiring the ultimate in
performance and functionality combined with speed of
operation and efficiency provided by the downloading of
data. When combined with software certification the
system provides a highly professional image for the
contractor combined with the ever increasing need for
tracability of test results.
- The BMM2580 offers all the features of the 2500 but with
the addition of 100 V and 50 V insulation ranges and a
10 uF capacitance range. These additional features make
the instrument ideally suited for applications in the
telecom industry.
- Both instruments offer comprehensive data storage options
with individual test results being stored against user
selectable distribution board and circuit references.
Alternatively data logging may be selected and up to 300
consecutive measurements of any chosen parameter may
be made at a selectable time interval. Data storage is
contained within non-volatile memory, (NVM), ensuring
that test results are not lost in the event of battery failure
or removal.
- Once stored, results may be recalled to the display,
printed to an external serial printer or downloaded to
suitable software. Download Manager™ for Windows is
included enabling language changes to be performed,
along with the simple download of stored data from the
instrument into comma separated variable files. These
files can be maintained as a record, used for manually
completing certificates or exported to other applications
such as spreadsheets and word processor documents.
Where creation of certificates of test is required both
instruments are fully compatible with PowerSuite for
Windows and NICEone. (for N.I.C.E.I.C. Certification
alone)
- Both instruments feature special mV ranges enabling
connection of a wide variety of mV output transducers.
Such devices extend the range of possible measurements
almost endlessly including such items as temperature
probes, airspeed indicators and high current clampmeters
thus extending the scope of the BMM2500 series into key
industries such as heating and ventilation (HVAC), and
servicing.
|
|
| |
| |
|
|
|
|
| |
|
| Insulation Ranges |
| Test Voltages : |
(BMM2500): 250, 500, 1000 V |
| (BMM2580): 50, 100, 250, 500, 1000 V |
| Measuring Range: |
(BMM 2500) 0,01 MΩ to 20 GΩ |
| (BMM 2580) 0,01 MΩ to 200 GΩ |
| (0 - 100 GΩ on analog scale ) |
| Test voltage accuracy: |
+15% maximum on open circuit |
| Test Current on load: |
1mA at min. pass value of
insulation specified in BS7671,
HD384 and IEC 364, 2 mA max.
|
| Accuracy: |
±2% ±2 digits ±0.2% per GΩ |
| Continuity |
| Measuring Range: |
0,01 Ω to 99,9 Ω (0 to 10 Ω on
analog scale) |
| Accuracy: |
±2% ±2 digits |
| Open circuit voltage: |
5 V ±1V |
| Test current: |
210 mA ±10 mA (0 - 2 Ω) |
| Buzzer: |
Operates continuously at less than
5 Ω |
| Resistance |
| Measuring Range: |
0,01 kΩ to 9,99 MΩ (0 to 100 MΩ on analog scale)
with Auto Ranging or Range Lock
Facility. |
| Accuracy: |
± 3% ± 2 digits |
| Open circuit voltage: |
5 V ±1V |
| Short circuit current |
25 μA ± 5 μA |
| Voltage |
| Measuring Range: |
±1V to ±500 V (0 to 1000 V on
analog scale) |
| Accuracy: |
0 to 500 V d.c. or a.c. (50/60 Hz)
±2% ±3 digit
0 to 500 V a.c. (400 Hz)
±5% ±3 digits |
| Millivolts |
| Measuring Range: |
±0,1 mV to 1 ±999 mV
(0 to 1000 mV on analog scale) |
| Accuracy: |
10 mV to 1999 mV d.c. or a.c
(50/60 Hz) ±2% ±3 digit
0,1 mV to 10 mV d.c. or a.c.
(50/60 Hz) ±2% ±5 digits
10 mV to 1999 mV a.c.
(16-460 Hz) ±5% ±3 digit
0,1 mV to 10 mV a.c.
(16-460 Hz) ±5% ±5 digits
d.c. millivolts zeroing: Up to
9,9 mV |
| Transducer Compatibility |
| Virtually any mV output transducer may be connected to facilitate
measurements of other parameters such as temperature, humidity
etc. A unique transducer offset adjustment ensures is included. |
| Capacitance (BMM2580) |
| Measuring Range: |
0,1 nF to 9,99 |
| Accuracy: |
±3% ±0,2 nF ±2 digits uF zeroing:
Up to 10 nF |
| Milliamperes |
| Measuring Range: |
0,1 mA to 500 mA (0 to 1000 mA
on analog scale) |
| Accuracy: |
10 mA to 1999 mA d.c. or a.c.
(50/60 Hz)
±2% ±3 digit
0,1 mA - 10 mA d.c. or a.c.
(50/60 Hz) ±2% ±5 digits
10 mA to 1999 mA a.c.
(16-460 Hz) ±5% ±3 digits
0,1 mA to 10 mA a.c.
(16-460 Hz) ±5% ±5 digits |
| Frequency |
| Measuring Range: |
16 Hz to 460 Hz |
| Accuracy: |
±1% ±1 digit |
| Data Logging |
| The instruments may be set to continuously record any one of the measurable parameters e.g. voltage, insulation resistance etc. The logging interval may also be adjusted to suit the application. |
| Logging Interval: |
10s to 1990s in 10s increments |
| Number of samples: |
Approximately 300 |
| Accuracy: |
As above for each parameter |
| Power Supply |
| Battery Type:: |
6 x 1,5 V Alkaline cells IEC LR6 type |
| Battery Life: |
Typically 3000, 5 second 1 kV tests |
| General |
| Backlight |
User selectable LED backlight with auto turn off to save battery
life |
| Test Result Storage |
Up to 99 Distribution Board References each with up to 99 Circuit
References may be specified. Test Results may be stored against
any circuit and distribution board combination.
Data may be recalled to the display, printed or downloaded to
PowerSuite for Windows or NICEone software |
| Communications |
RS232 9 pin male D connector.
Baud rate 9600 |
| Printer Output |
Prints test results, distribution board and circuit details to external
serial print |
|
|
| |
| |
|
|
|
|
| |
| |
 |
|
 |
|
|
 |
 |
|
|
|
|
Safety
The BMM2500 series
complies with the latest international
directives concerning safety and electromagnetic compatibility
The instruments meet the requirements for double insulation to
IEC 61010-1 (1995), EN 61010-1 (1995) Safety Requirements for
electrical equipment for measurement, control, and laboratory
use.
Category III**, 300 Volts phase to earth (ground) and 440
Volts phase to phase, without the need for separately fused test
leads. If required, fused test leads are available as an optional
accessory.
** Relates to the transient over-voltages likely to be met in fixed
wiring installations.
Complies with the following parts of EN 61557, Electrical safety
in low voltage systems up to 1000 V a.c. and 1500 V d.c.
Fuse:
500 mA (F) 500 V, 32 x 6 mm Ceramic HBC 10 kA minimum.
Environmental
Conditions
Operating range:
-5 to +40ºC
Operating humidity:
90% RH at 40ºC max.
Storage
Temperature range:
-25 to +65ºC
Calibration Temperature: +20ºC
Maximum altitude
: 2000 m
Dust and water protection: IP54
Temperature coefficient:
<0,1% per ºC
Physical Specifications
Dimensions
Length:
220 mm (8.66 inches)
Depth:
110 mm (4.33 inches)
Height: 45 mm (1.77 inches)
Weight:
742 g (1.63lbs)
(including batteries) |
|
| |
| |
|
|