Hsiang Cheng Model FPTR

RTD Isolating Industrial Transducer Series

space

Hsiang Cheng FPTR, RTD Isolating Transducers

Product Parameters:
space
Input modulesMax. effective ranges Input Ranges Input Ranges OutputRanges Power
-100~110C / -170~170F-100~440C / -150~800F-100~800C / -150~1500F-100~250C / -150~450FNote : Module P1/P4/P8 for PT 100Ω Module C4 for CU 10Ω -100~0-100~50-50~50-50~100 -50~200 0~100 0~150 0~200 0~250 0~300 0~400 0~500 0~600 0~700 0~800 0~9000~10000~11000~1200 0~1400 0~1500 Specified 0~1 V0~2 V0~5 V1~5 V 0~10 V 2~10 V 0~1 mA 0~2 mA 0~5 mA 1~5 mA 0~10 mA 2~10 mA 0~16 mA 0~20 mA 4~20 mA AC120VAC240VOptionDC24V DC48V DC12V DC125V
Model Type Description
FPTR-1 C FPTR-1 C single otput custom calibration – °C
FPTR-2 C FPTR-2 C dual output custom calibration – °C
FPTR-1 F FPTR-1 F single otput custom calibration – °F
FPTR-2 F FPTR-2 F dual output custom calibration – °F
space
Specifications :
space
General Specifications
Accuracy ( 23±3°C ) 0.3% ro
Linearty & repeatability 0.1% typical
Dielectric strength 4KV rms AC between power / input & output terminals
Surge and impulse test ANSI C37.90 / 1989, IEC 255-3 (1989) 4 KV 1.2 x 50 us
Configuration Three wires connection / applicable for two wiresExcitation – 5mAdc constant current sourceChangeable input moduleLinearized output vs input temperature
Field rangeability Selectable output mode for voltage or current modeSwitchable of input range & output rangeSwitchable AC power 110V or 220VSelectable degree C or F calibration
Effect of wire resistance Automatically eliminated for 3 wires connectionOffset recalibration required as for 2 wires connection
Input break detection Selectable Hi-set or Lo-setHi-set over 110% of rated outputLo-set less 0.5% of rated output
Aux. power effect < 0.003% for per voltage change
Storage condition Temperature range -25 to 70°C, RH 20 to 95% non condensed
Operating condition Temperature range -20 to 55 °C, RH 0 to 99% non condensed
space
pdf_pic Downloads: General Specifications