- TRMS measurements when connected to a TRMS instrument
- No core saturatin or damage if overloaded
- Waterproof sensor
- 9V Alkaline battery for typical use of 2000 hrs continuous operation
- Shape Memory for pre-shaping of sensor before use (no drooping)
- Insensitive to DC, measures only AC component on DC + AC
- Excellent linearity
- EN 61010, 1000V Cat. III (sensor); EN 61010, 600V Cat. III (module)
|
|
| |
| |
|
|
|
| |
| |
- Load analysis
- Data logging
- power quality monitoring
- General purpose current monitoring
|
|
| |
| |
|
|
|
| |
 |
The LongLife AmpFlex® Model LGLF 3000-36-2-1 is a flexible AC current probe composed of a flexible sensor and an electronic module. The flexible sensor permits measurements on conductors where standard clamp-on probes cannot be used. The 9V Alkaline battery provides typical use of 2000 hours of continuous operation. Dual measurement range is 5 to 300A and 50 to 3000A. Output signal is 3.33mV/A and 0.333mV/A. Sensor length is 36". |
|
| |
| |
|
|
|
| |
| |
| LGLF 3000-36-1-1 |
|
| ELECTRICAL |
|
| Current Range |
3000A |
| Measurement Range |
5 to 300A
50 to 3000A |
| Output Signal(s) |
3.33mV/A &
0.333mV/A |
| Residual Current |
200mArms |
| Overload |
500A max |
| Phase Shift |
<1° @ 50 to 60Hz |
| GENERAL |
| Accuracy |
1% of Reading + 200mA
from 50 to 3000A
1% of Reading + 500mA
from 5 to 50A |
| Weight |
0.8 lb (13 oz) |
| Maximum Conductor |
24": Ø 7.64" (194mm); 36": Ø 11.46" (291mm) |
| Size |
6.5 ft (2m) sensor to cast |
| Cable Length
Output |
14" (0.35m) double insulated cable with 4mm banana plug |
| MECHANICAL |
| Safety Standards |
Sensor: 24" or 36" (609 or 914mm)
Module: 1.56 x 2.5 x 4.5" (40 x 64 x 114mm) |
| Operating Temperature and %RH |
14° to 131°F (-10° to 55°C) (sensor to 90°C),
10 to 90% RH (@50°C) |
| Minimum Bend Radius |
0.75" (19mm) |
| ENVIRONMENTAL |
| Operating Temperature |
14° to 131°F (-10° to 55°C) |
| Storage Temperature |
-4° to 158°F (-20° to 70°C) |
Operating Relative
Humidity |
50° to 86°F (10° to 30°C) (non-condensing)
104° to 122°F (40° to 50°C) (non-condensing) |
| SAFETY |
| Rating |
EN 61010, 1000V, Cat. III, Pollution Degree: 2 |
|
|
| |
| |
|
|
|
| |
| |
Specifications |
|
|
|
 |
| |
 |
|
|
.jpg) |
|
The LongLife AmpFlex® is a flexible
AC current probe composed of a
flexible sensor and an electronic
module. The flexible sensor permits
measurements on conductors where
standard clamp-on probes could not
be used. In particular, it can be installed
in tight spaces, around breaker panels,
around cable bundles, around wide or
large bus bars, or even wrapped
around irregular shapes. The Shape
Memory feature enables the user to
“pre-shape” the sensor, somewhat,
before inserting it between or around
conductors. This feature facilitates
closing, enhances user safety, and
alleviates the drooping effect typically
associated with flexible sensors.
The LongLife AmpFlex® is lightweight.
It does not use magnetic cores like
standard transformers. The transformation
principle is based on an air
core. It presents virtually no load to
the system under test, has a low
phase shift and excellent frequency
response, and cannot be damaged
by overloads. The sensor assembly
is waterproof and insulated for
1000V Cat. III.
The LongLife AmpFlex® has a mV
output proportional to the current
measured for direct readings on DMMs,
data loggers and oscilloscopes. TRMS
measurements are taken when
connected to a TRMS meter. The
LongLife AmpFlex® is insensitive to DC
currents and only the AC component
of the measured signal is measured. |