.jpg) |
|
|
|
| |
|
| |
 |
- Compact size - fits into your pocket
- 400 AAC current measurements
- 600VAC/DC volts measurements
- Resistance measurements to 400Ω
- Continuity with beeper below 40Ω
- Frequency measurements from V and A
- Hold function to "freeze" readings
- Large easy-to-read 4000-count display
- 42-segment bargraph
- Includes test leads and a soft carrying pouch
|
|
|
|
| |
| |
|
|
|
|
| |
| |
- Industrial, commercial, HVAC and residential troubleshooting
- Power panel, junction box and battery bank monitoring
- Electrical testing and troubleshooting on traditional and non-linear leads, such as adjustable speed drives and computers
|
|
| |
| |
|
|
|
| |
 |
The Model 500, Clamp-on Meter, is a general-purpose professional Clamp-on. A strong mechanical case design, quality materials, full ranges and compliance to international safety and quality standards ensure a professional and reliable tool. The Model 501 is auto-ranging, and provides the user with AC Volts, DC Volts, AC Amperes, resistance, and continuity. The Model 501 also includes a frequency function.
The large and complete 4000-count LCD features comprehensive user interface symbols such as a low battery indicator. It is equipped with a Data Hold function. The Model 501 is compact for comfortable one-handed operation, and the hooked jaw design is easy to maneuver in crowded wiring and breaker panels. The jaw opening for the Model 501 is 1.1" (28mm). |
|
| |
| |
|
|
|
| |
| |
| Average Sensing |
Model 501 |
| AC AMPERES |
|
| Measurement Range |
0.05 to 400A |
| Resolution |
40A
400A
|
0.01A
0.1A |
| Accuracy |
50 to 60Hz
60 to 500Hz |
±1.9% of Reading ± 5cts
±2.5% of Reading ± 5cts |
| Overload Protection |
600Arms |
| AC VOLTS |
(Auto-Ranging) |
| Measurement Range |
0.5 to 600V |
| Resolution |
400V
600V |
0.1V
1V |
| Accuracy |
50 to 60Hz
60 to 500Hz |
±0.8% of Reading ± 5cts
±1.5% of Reading ± 5cts |
| Input Impedance |
1MΩ |
| Overload Protection |
600Vrms |
| DC VOLTS |
(Auto-Ranging) |
| Measurement Range |
0.2 to 600V |
| Resolution |
400V
600V |
0.1V
1V |
| Accuracy |
±1% of Reading ± 2cts |
| Input Impedance |
1MΩ |
| Overload Protection |
600Vrms
|
| RESISTANCE-OHMS |
(Ω) |
| Measurement Range |
0.2 to 400Ω |
| Resolution |
0.1Ω |
| Accuracy |
±1% of Reading ± 2cts |
| Max Test Voltage |
1.5VDC |
| Overload Protection |
600Vrms
|
| CONTINUITY |
|
| Beeper Activation |
<40Ω |
| Max. Test Voltage |
1.5VDC |
| Overload Protection |
600Vrms |
| FREQUENCY (Hz) |
(Auto-Ranging) |
| CURRENT INPUT |
|
| Range |
|
20Hz - 4kHz
>4kHz - 10kHz |
1Hz Resolution
10Hz Resolution |
| Accuracy |
±0.1% of Reading ± 1ct |
| Min Input Signal |
2Arms |
| Input Impedance |
1MΩ |
| Overload Protection |
600Arms |
| VOLTAGE INPUT |
|
| Range |
|
10Hz to 4kHz
>4kHz to 40kHz
>40kHz to 400kHz
|
1Hz Resolution
10Hz Resolution
100Hz Resolution
|
| Accuracy |
±0.1% of Reading ± 1ct
|
| Min Input Signal |
|
4kHz
40kHz
400kHz
|
5Vrms
5Vrms
5Vrms
|
| Input Impedance |
1MΩ |
Overload Protection |
600Vrms |
|
|
| |
| General Specifications |
| Digital Display |
3 3/4 digits LCD display, 4000-count (max. reading 3999) |
| Analog Display |
Fast 42-segment analog bargraph display |
| Symbol & Scale Range |
Automatic according to range and input signal |
| Polarity |
Displayed when negative signal applied to input |
| Overload |
Displayed when input signal exceeds range |
| Sample Rate |
2 samples/sec for the digital display, 20 samples/sec for the analog bargraph |
| Power Source |
Two 1.5V AAA batteries |
| Auto-Off |
30 minutes |
| Low Power Indication |
Displayed when the battery is below the required voltage |
| Jaw Opening Size Ø |
1.1" (28mm) – faciliates 500MCM cable |
| Dimension |
7.60 x 1.97 x 1.1" (193 x 50 x 28mm) |
| Weight |
8.11 oz (230g) with batteries |
| ENVIRONMENTAL |
|
| Altitude |
2167 ft (2000m) max. |
| Operating Temperature |
32° to 104°F (0° to 40°C),
<80% RH, non-condensing |
| Storage Temperature |
14° to 140°F (-10° to 60°C),
<70% RH, battery removed |
| Pollution Degree |
2 |
| SAFETY |
|
| Safety Rating |
EN 61010, 600V Cat. II, EN 61010, 300V Cat. III |
| Double Insulation |
Yes |
| CE Mark |
Yes |
|
| |
| |
|
|
|
| |
| |
Specifications |
|
|
|
 |
| |
 |
| |
Related urls: |
|
|
| |
- Model 500 , (AC, 400AAC, 600VAC, Ohms, Continuity)
- Model 501 , (AC, 400AAC, 600VAC/DC, Ohms, Continuity)
- Model 502 , (TRMS, 400AAC, 600VAC/DC, Ohms, Continuity)
- Model 503 , (AC/DC, 400AAC/DC, 600VAC/DC, Ohms, Continutiy)
- Model 511 , (AC, 1000AAC, 1000VAC/DC, Ohms, Continuity)
- Model 512 , (TRMS, 1000AAC, 600VAC/DC, Ohms, Continuity)
- Model 514 , (AC/DC, TRMS, 1000AAC/DC, 600VAC/DC, Ohms,
Continuity)
- Model F1N ,(AC, TRMS, 400A, Hz)
- Model F3N , (AC, TRMS, 400A, Hz, Peak, Smooth, Min/Max)
- Model F01 , (TRMS, 400AAC, Ohms, Continuity)
- Model F03 , (TRMS, 400AAC/DC, 600VAC/DC, Ohms, Continuity, Temp.)
- Model F05 , (TRMS, 400AAC/DC, 600VAC/DC, Hz, Power (kW), Phase Rotation, Ohms, Continuity)
- Model F07, (TRMS, 400AAC/DC, 600VAC/DC, Hz, Ohms, Continuity, Temperature)
- Model F25 , (TRMS, 600VAC/DC & 1500AAC/DC)
- Model 721 , (9999-count, TRMS, V, A, F, Hz, THD, CF)
|
|
| |
|
| |
|
|
|